Skip to content

Arthrex® AR-7000-13Recalls & reported problems

Arthrex, Inc. · CANNULATED HEX DRIVER, 2.5MM X 7" LONG · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867040892), received between Jan 3, 2024 and Jan 3, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Erosion” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Erosion1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 3, 2024Material Erosion; Material Erosion

    ON THE 8TH DECEMBER 2023, IT WAS REPORTED BY A ARTHREX REP. VIA EMAIL THAT FOR AN AR-7000-13, (CANNULATED HEX DRIVER, 2.5 MM X 7 '' LONG) THE SCREWDRIVER SHAFT WAS FOUND TO BE TOO STRIPPED TO ENGAGE PROPERLY WITH THE 3.75 LATARJET SCREWS. THIS WAS DETECTED DURING USE IN A LATARJET PROCEDURE ON 5TH DECEMBER 2023. THE CASE WAS COMPLETED SUCCESSFULLY WITH A NON-ARTHREX SCREWDRIVER TO COMPLETE INSERTION OF THE SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18437219 ↗

Other INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices

All INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page