Skip to content

Arthrex® AR-8150PP-45Recalls & reported problems

Arthrex, Inc. · POWERPICK, 45 DEG · Class 1

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867042919), received between Sep 21, 2021 and Sep 13, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Material Integrity Problem1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 13, 2024Break; Break

    ON (B)(6) 2024 IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA SEMS-06649114 THAT AN AR-8150PP-45 POWERPICK WHEN THE TIP OF THE PRODUCT WAS PRESSED AGAINST THE AFFECTED AREA, THE TIP SHRANK. THIS WAS DISCOVERED DURING THE CASE. THIS CASE WAS COMPLETED BY USING ANOTHER PRODUCT OF SAME PRODUCT NUMBER WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20213294 ↗
  • MalfunctionAug 15, 2024Break; Break

    IT WAS REPORTED THAT DRILL TIP AND THE SHAFT BROKE DURING A DEMONSTRATION. IT WAS NOTICED BEFORE THE SURGERY. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THERE WAS NO CASE INVOLVEMENT REPORTED. NO FURTHER INFORMATION RECEIVED. UPDATE DW 18-JUL-2024: FURTHER INFORMATION WAS RECEIVED THAT THE REPORTED EVENT OCCURRED DURING A HIP-ARTHROSCOPY. UPDATE DW 30-JUL-2024: THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER AND ALL BROKEN FRAGMENTS WERE RETRIEVED OUT OF THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19992616 ↗
  • MalfunctionJul 18, 2022Break

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY SURGERY THE HANDLES OF THE FOUR DEVICES BROKE. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE FROM ANOTHER MANUFACTURER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE 30-JUN-2022 IT WAS CONFIRMED THAT THE PLASTIC HANDLE BROKE, WHICH PREVENTED THE WIRE FROM SLIDING OUT. ALL BROKEN PARTS WERE RETRIEVED FROM THE PATIENT. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15040488 ↗
  • MalfunctionSep 21, 2021Mechanical Problem; Material Integrity Problem

    IT WAS REPORTED THAT DURING A ARTHROSCOPIC PROCEDURE WHILE DRILLING IN THE JOINT DEBRIS WAS PRODUCED. THE SURGEON WAS ABLE TO RETRIEVE EVERYTHING OT OF THE PATIENT. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12506256 ↗

Other BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL devices

All BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL devices →