Skip to content

Arthrex® AR-8300Recalls & reported problems

Arthrex, Inc. · APSII CONTROL CONSOLE · Class 1

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867043138), received between Sep 8, 2023 and Feb 14, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Connection Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Connection Problem1
  • Defective Device1
  • Electromagnetic Interference1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 14, 2024Electromagnetic Interference; Electromagnetic Interference

    IT WAS REPORTED THAT THE SHAVER WAS USED DURING SURGERY WHEN PATIENT GOT CARDIAC ARRHYTHMIA. SURGEON STOPPED THE SHAVER. WHEN THE SURGEON USED THE SHAVER AGAIN, THE CARDIAC ARRHYTHMIA OCCURRED AGAIN WITH THE PATIENT. AFTER SECOND USE THE SURGEON WAS FINISHED WITH USAGE OF THE SHAVER AND THEREFORE HE DIDN´T NEED TO CHANGE THE SHAVER. THE PATIENT IS WELL. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. ***UPDATE SWIT 15-DEC-2020: THE ARTHREX SHAVER AR-8330F ONLY INFLUENCED THE ECG DEVICE IN THE DISPLAY OF CONDUCTION. THE PATIENT DID NOT HAVE ANY ARRHYTHMIAS AT ANY TIME. THE SHAVER IN QUESTION WAS REPLACED DURING THE SURGERY. WITH THE REPLACEMENT DEVICE, THERE WAS NO FURTHER CHANGE ON THE ECG MONITOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18703689 ↗
  • MalfunctionOct 30, 2023Defective Device; Defective Device

    IT WAS REPORTED THAT THE SHAVER CONSOLE IS DEFECTIVE. IT DOES NOT WORK WHEN PLUGGING IN THE SHAVER HANDPIECE. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18031484 ↗
  • MalfunctionSep 8, 2023Connection Problem; Connection Problem

    IT WAS REPORTED THAT WITH THE ATTACHED HANDPIECE THE SCREEN SHOWES THAT IT IS NOT CONNECTED. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. NO FURTHER INFORMATION RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17709945 ↗

Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices

All INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.