Arthrex® AR-8400BCRecalls & reported problems
Arthrex, Inc. · BONE CUTTER, 4.0MM X 13CM · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00888867043466), received between May 3, 2018 and Sep 16, 2026 — classified as 2 injuries, 16 malfunctions.
The most frequently reported specific device problem is “Physical Resistance/Sticking” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Physical Resistance/Sticking5
- Break4
- Material Fragmentation4
- Defective Device3
- Material Twisted/Bent2
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- Insufficient Information5
- Injury1
Recent adverse event reports
The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2026Material Fragmentation
Patient problems: Insufficient Information
MAUDE report 50788481 ↗MalfunctionSep 16, 2026Material Fragmentation
Patient problems: Insufficient Information
MAUDE report 50787101 ↗MalfunctionDec 29, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
ON 8-DEC-2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-8400BC BONE CUTTER IS STICKING AT THE MOUTH AND NOT OSCILLATING. NOTICED DURING A CASE, SWAPPED, AND CASE COMPLETED WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23906502 ↗MalfunctionDec 29, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
ON 8-DEC-2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-8400BC BONE CUTTER IS STICKING AT THE MOUTH AND NOT OSCILLATING. NOTICED DURING A CASE, SWAPPED, AND CASE COMPLETED WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23906490 ↗MalfunctionDec 29, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
ON 8-DEC-2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-8400BC BONE CUTTER IS STICKING AT THE MOUTH AND NOT OSCILLATING. NOTICED DURING A CASE, SWAPPED, AND CASE COMPLETED WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23906489 ↗MalfunctionDec 29, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
ON 8-DEC-2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-8400BC BONE CUTTER IS STICKING AT THE MOUTH AND NOT OSCILLATING. NOTICED DURING A CASE, SWAPPED, AND CASE COMPLETED WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23906488 ↗MalfunctionDec 29, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
ON (B)(6) 2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-8400BC BONE CUTTER IS STICKING AT THE MOUTH AND NOT OSCILLATING. NOTICED DURING A CASE, SWAPPED, AND CASE COMPLETED WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23906485 ↗MalfunctionDec 24, 2025Break; Break
IT WAS REPORTED THAT DURING THE ANKLE ARTHROSCOPY, THE DEVICE BROKE INSIDE THE JOINT. THE BROKEN PIECES WERE RETRIEVED FROM THE PATIENT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23888628 ↗MalfunctionFeb 5, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT DURING A BROSTROM SURGERY THE SHAVER SHEDS METAL CHIPS IN THE ANKLE JOINT. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE (B)(6) 16-JAN-2025: FURTHER INFORMATION WAS PROVIDED THAT THE PIECES WERE RETRIEVED OUT OF THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21311379 ↗InjuryJun 21, 2024Break; Break
UPDATE 23-JUL-2024: DURING A PHONE CALL THE FOLLOWING ADDITIONAL INFORMATION WERE MADE AVAILABLE. IT WAS FURTHER CONFIRMED THAT ALL DEVICES AR-8500OBE, AR-8400BC AND AR-6420CEX FAILED DURING THE SAME SURGERY, BECAUSE THE SURGEON HAD TO INCREASE THE PRESSURE AS THE PATIENT BONE WAS VERY HARD.
Patient problems: Insufficient Information
MAUDE report 19588688 ↗MalfunctionJun 19, 2024Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING A PROCEDURE, THE AR-8400BC BENT. THE CASE WAS COMPLETED BY USING A NEW AR-8400BC.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19566575 ↗MalfunctionFeb 7, 2024Defective Device; Defective Device
ON (B)(6) 2024, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA EMAIL THAT (2) AR-8400BC BONE CUTTER BECAME DAMAGED DURING A PROCEDURE ON (B)(6) /2024. NO FURTHER INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION REQUESTED.
Patient problems: Insufficient Information
MAUDE report 18661375 ↗MalfunctionFeb 7, 2024Defective Device; Defective Device
ON (B)(6) 2024, IT WAS REPORTED BY AN ARTHREX SUBSIDIARY EMPLOYEE VIA EMAIL THAT (2) AR-8400BC BONE CUTTER BECAME DAMAGED DURING A PROCEDURE ON (B)(6) 2024. NO FURTHER INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION REQUESTED.
Patient problems: Insufficient Information
MAUDE report 18661365 ↗MalfunctionJan 26, 2024Material Fragmentation; Material Fragmentation
ON (B)(6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS (B)(4) THAT, DURING A CASE, THE BONE CUTTER (AR-8400BC) WAS PRODUCING METAL SHAVINGS INTO THE JOINT. THE DEVICE WAS REPLACED AND THE CASE WAS COMPLETED WITH NO ADVERSE EFFECT TO THE PATIENT. ADDITIONAL INFORMATION REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18586274 ↗MalfunctionOct 25, 2023Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING AN ARTHROSCOPY SURGERY THE TIP OF THE DEVICE GOT BENT. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18004384 ↗MalfunctionJul 18, 2023Defective Device; Defective Device
ON (B)(6) 2023, IT WAS REPORTED BY A ARTHREX EMPLOYEE VIA SEMS THAT AN AR-8400BC BONE CUTTER AND AR-8400PR POWERRASP DO NOT WORK. THE SAME PROBLEM CONTINUES 2 TIMES. THIS WAS DISCOVERED DURING A CASE, USED A NEW PRODUCT AND THE CASE IS COMPLETED. NO PATIENT HARMS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17343499 ↗MalfunctionSep 19, 2022Break
ON (B)(6) 2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT A AR-8400BC BONE CUTTER BROKE. THIS OCCURRED ON (B)(6) 2022 DURING A SUB-ACROMIAL DECOMPRESSION OF THE SHOULDER WHEN THE SHAFT OF THE DEVICE BROKE IN HALF. THE PATIENT HAD GOOD BONE QUALITY, AND THE BROKEN FRAGMENT WAS RETRIEVED. THE CASE WAS COMPLETED USING ANOTHER BONE CUTTER. THERE WAS NO PATIENT EFFECT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15445804 ↗InjuryMay 3, 2018Break
IT WAS REPORTED THAT THE BONE CUTTER BROKE IN THE HUB DURING SURGERY. ALL BROKEN PIECES WERE RETRIEVED FROM THE PATIENT. HOWEVER, THE PATIENT WAS BURNT. OPENED ANOTHER BONE CUTTER AND COMPLETED THE PROCEDURE. SHOULDER SCOPE. FOLLOW UP INFORMATION: SALES REP WAS NOTIFIED ABOUT THE INCIDENT ON (B)(4) 2013. IT HAS BEEN REPORTED THAT THE ACCOUNT/NURSING DIRECTOR DID NOT KNOW THE MEANS OF HOW THE PATIENT WAS BURNT. ACCOUNT REPORTED TO THE SALES REP THAT THE BURN WAS LOCATED AT ONE OF THE INCISIONS PORTALS ON THE ANTERIOR ASPECT OF THE SHOULDER. IT WAS REPORTED TO BE A MINOR SUPERFICIAL BURN WHICH WAS TREATED PER FACILITY POLICY. PATIENT WAS DISCHARGED SAME DAY.
Patient problems: Injury
MAUDE report 7483540 ↗
FDA clearance and classification
The FDA regulates product code GFA (blade, saw, general & plastic surgery, surgical) under 21 CFR 878.4820 as a Class I device. 21 CFR 878.4820 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Arthrex® AR-8400BC
Has the Arthrex® AR-8400BC been recalled?
No FDA recall record in our data names this exact device identifier (DI 00888867043466), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Arthrex® AR-8400BC?
The FDA's MAUDE database lists 18 adverse event reports for this device, received between May 3, 2018 and Sep 16, 2026. In those reports, the most common specific device problems are “Physical Resistance/Sticking” (5), “Break” (4) and “Material Fragmentation” (4); the most common patient problem is “Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Arthrex® AR-8400BC?
None of the reports are classified as deaths, 2 as injuries and 16 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Arthrex® AR-8400BC FDA approved?
Devices of this type (product code GFA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Arthrex, Inc.
Who makes the Arthrex® AR-8400BC, and is it still sold?
The FDA's device database lists Arthrex, Inc. as the company responsible for it (brand Arthrex®, model AR-8400BC). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL devices
- NA 511.801Synthes GmbH · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL223 reports
- Formula 0375544000Stryker Corporation · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL98 reports
- NA 0277096325Stryker Corporation · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL95 reports
- NA 310.25Synthes GmbH · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL91 reports
- Arthrex® AR-8400TDArthrex, Inc. · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL86 reports
- NA 0277096275Stryker Corporation · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL72 reports
- Arthrex® AR-7300DSArthrex, Inc. · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL71 reports
- NA 0277096277Stryker Corporation · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL69 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page