Skip to content

Arthrex® AR-8400PRRecalls & reported problems

Arthrex, Inc. · POWERASP,4.0MM X 13CM · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00888867043619), received between Aug 31, 2021 and Aug 13, 2024 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 3 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Melted3
  • Defective Device1
  • Device Slipped1
  • Excessive Heating1
  • Failure to Cut1
  • Loose or Intermittent Connection1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Insufficient Information2

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 13, 2024Melted; Melted

    ON 08/07/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS-(B)(4) THAT (2) AR-8400PR POWERASP CAME APART. THIS OCCURRED DURING USE THE HOUSING THAT INTERACTS WITH THE SHAVER HANDPIECE MELTED AND DISSEMBLED. THE UNITS FUNCTIONED FINE LONG ENOUGH FOR THE SURGEON TO COMPLETE THE CASE, HOWEVER, THEY THEN STOPPED WORKING AND CAME/MELTED APART.

    Patient problems: Insufficient Information

    MAUDE report 19974609 ↗
  • MalfunctionJun 7, 2024Overheating of Device; Overheating of Device

    IT WAS REPORTED ON 03/24/2022 BY A FACILITY REPRESENTATIVE VIA (B)(4) THAT AN AR-8400PR POWEERRASP OVERHEATED AND BROKE. THIS OCCURRED DURING A CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19491799 ↗
  • MalfunctionJan 3, 2024Melted; Melted

    IT WAS REPORTED THAT DURING AN ARTHROSCOPIC SURGERY THE HUB OF THE DEVICE MELTED. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18435769 ↗
  • MalfunctionDec 3, 2023Break; Break

    IT WAS REPORTED THAT THE GRATER BROKE DURING A SURGERY. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18253509 ↗
  • MalfunctionDec 3, 2023Break; Break

    IT WAS REPORTED THAT THE GRATER BROKE DURING A SURGERY. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18253500 ↗
  • MalfunctionDec 3, 2023Break; Break

    IT WAS REPORTED THAT THE GRATER BROKE DURING A SURGERY. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18253498 ↗
  • MalfunctionAug 4, 2023Melted; Melted

    IT WAS REPORTED THAT DURING AN ANKLE FUSION SURGERY THE POWERASP MELTED AND CAME APART IN THE HAND PIECE. THE SAME THING HAPPENED WITH SECOND ONE. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH DIFFERENT PART NUMBER. IT WAS NOT NECESSARY TO DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17464679 ↗
  • MalfunctionAug 4, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT DURING AN ANKLE FUSION SURGERY THE POWERASP MELTED AND CAME APART IN THE HAND PIECE. THE SAME THING HAPPENED WITH SECOND ONE. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH DIFFERENT PART NUMBER. IT WAS NOT NECESSARY TO DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17461978 ↗
  • MalfunctionJul 18, 2023Defective Device; Defective Device

    ON (B)(6) 2023, IT WAS REPORTED BY A ARTHREX EMPLOYEE VIA SEMS THAT AN AR-8400BC BONE CUTTER AND AR-8400PR POWERRASP DO NOT WORK. THE SAME PROBLEM CONTINUES 2 TIMES. THIS WAS DISCOVERED DURING A CASE, USED A NEW PRODUCT AND THE CASE IS COMPLETED. NO PATIENT HARMS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17343332 ↗
  • MalfunctionSep 14, 2021Thermal Decomposition of Device; Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT DURING A SHOULDER ARTHROSCOPY THE SHAVER ATTACHMENT IS SCORCHED / BURNT THROUGH. NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY WAS REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12466757 ↗
  • MalfunctionAug 31, 2021Loose or Intermittent Connection; Device Slipped; Failure to Cut

    IT WAS REPORTED THAT DURING AN SCR PROCEDURE PERFORMED ON AN (B)(6) YEAR-OLD FEMALE, THE POWERASP WAS UNABLE TO CUT BONE FROM THE START OF THE SURGERY. THE SWIVELOCK FELL OFF AND THE INSERTER SHAFT WAS LOOSE AND NOT FIXED; THE ISSUE DID NOT RESOLVE WHEN IT WAS RETURNED TO THE BASE OF THE INSERTER. THE PROCEDURE WAS COMPLETED USING A NEW PRODUCT WITH NO PATIENT HARM REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 12393511 ↗

Other BUR, SURGICAL, GENERAL & PLASTIC SURGERY devices