Skip to content

Arthrex® AR-8500COERecalls & reported problems

Arthrex, Inc. · BUR,OVAL,CLEARCUT,8 FLUT,5.0MMX 13CM · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867043756), received between Feb 3, 2021 and Sep 28, 2023 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Break1

Patient problems

  • Foreign Body In Patient1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A ROTATOR CUFF SURGERY THE LITTLE PLASTIC/ACRYLIC PIECE DETACHED FROM THE BURR AND STAYED INSIDE OF THE PATIENT. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: Foreign Body In Patient

    MAUDE report 17839744 ↗
  • MalfunctionJul 18, 2022Detachment of Device or Device Component

    ON 06/29/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT ON (B)(6) 2022 DURING A ROTATOR CUFF REPAIR A AR-8500COE BURRS CLEAR PORTION ON THE TIP CAME OFF INTO THE PATIENT. THE FRAGMENT WAS RETRIEVED, AND THE CASE WAS COMPLETED USING A SMALLER BURR. THERE WAS NO PATIENT EFFECT REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 15040485 ↗
  • MalfunctionFeb 3, 2021Break

    IT WAS REPORTED THAT THE HOOD OF THE AR-8500COE BROKE OFF DURING USE FOR A RCR WITH A DISTAL CLAVICLE EXCISION AND SUB-ACROMINAL DECOMPRESSION. THE PIECE WAS RETRIEVED FROM THE PATIENT; MALE, (B)(6) YEARS OLD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11274345 ↗

FDA clearance and classification

The FDA regulates product code GFF (bur, surgical, general & plastic surgery) under 21 CFR 878.4820 as a Class I device. 21 CFR 878.4820 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Arthrex® AR-8500COE

Has the Arthrex® AR-8500COE been recalled?

No FDA recall record in our data names this exact device identifier (DI 00888867043756), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Arthrex® AR-8500COE?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Feb 3, 2021 and Sep 28, 2023. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (2) and “Break” (1); the most common patient problem is “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Arthrex® AR-8500COE?

None of the reports are classified as deaths, 0 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Arthrex® AR-8500COE FDA approved?

Devices of this type (product code GFF) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Arthrex, Inc.

Who makes the Arthrex® AR-8500COE, and is it still sold?

The FDA's device database lists Arthrex, Inc. as the company responsible for it (brand Arthrex®, model AR-8500COE). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other BUR, SURGICAL, GENERAL & PLASTIC SURGERY devices

All BUR, SURGICAL, GENERAL & PLASTIC SURGERY devices →

Sources and updates