Skip to content

Arthrex® AR-8943-12Recalls & reported problems

Arthrex, Inc. · DRV SFT,CANN,QC,T15 HEXALOBE · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867052222), received between May 10, 2021 and May 22, 2025 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Defective Device1
  • Device Damaged by Another Device1
  • Flaked1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 22, 2025Break; Break

    FURTHER INFORMATION WAS RECEIVED THAT A SMALL SURGERY MEANS AN ANKLE SURGERY IN THIS CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22074234 ↗
  • InjurySep 27, 2023Defective Device; Device Damaged by Another Device; Defective Device; Device Damaged by Another Device

    ON 30TH AUGUST 2023, IT WAS REPORTED BY A SALES REP VIA EMAIL CONCERNING 2 PIECES OF AR-8840C-45, LOW PROFILE SCREW¿, SS, 4.0 X 45 MM, CANNULATED, SHORT THREAD. FOR 1 OF THE AR-8840C-45, ITS SCREW HEAD WAS STRIPPED. BOTH WERE EVENTUALLY REMOVED FROM THE PATIENT DUE TO REDUCTION LOSS. THIS OCCURRED DURING A LEFT OPEN REDUCTION INTERNAL FIXATION ANKLE FRACTURE PROCEDURE ON (B)(6) 2023. AS PART OF A LEFT ANKLE ORIF (ALSO INCLUDING PLATE AND SYNDESMOSIS TIGHTROPES), SURGEON ATTEMPTED TO INSERT 2X 4.0MMX 45MM SHORT-THREAD CANNULATED SCREWS AR-8840C-45 INTO THE MEDIAL SIDE OF THE ANKLE. PER TECHNIQUE, THE SURGEON FIRST USED 1.35MM GUIDE WIRE AR-8943-01, GUIDE WIRE, TROCAR TIP, 1.35 MM X 130 MM, THEN OVERDRILLED WITH A AR-8943-02, DRILL BIT, CANNULATED, 2.6 MM. WHEN INSERTING THE SCREW USING THE AR-8943-12, DRIVE SHAFT, CANNULATED, QC, T15 HEXALOBE PER SURGICAL TECHNIQUE, THE SURGEON STATED THAT SHE WAS UNABLE TO ADVANCE THE SCREW PROPERLY AS THE DRIVER SHAFT WAS NOT ENGAGING PROPERLY WITH THE SCREW. FOR THE SECOND IDENTICAL SCREW, THE SURGEON SWITCHED TO THE AR-8943-09, SCREWDRIVER, T15 HEXALOBE, CANNULATED AND THE SURGEON STATED THAT THIS SCREW ADVANCED MORE SUCCESSFULLY. UPON IMAGING,

    Patient problems: Insufficient Information

    MAUDE report 17824949 ↗
  • MalfunctionMay 10, 2021Flaked

    IT WAS REPORTED THAT DURING A PROCEDURE, THE PAINT WAS FLAKING OFF OF THE FOLLOWING DEVICES. THE FLAKES WERE REMOVED FROM THE PATIENT. NO PATIENT HARM. AR-8943-09, LOT: 1391715, AR-8943-12, LOT: 6791217, AR-8943-14, LOT: 5262019.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11799554 ↗

Other INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices

All INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page