Skip to content

Arthrex® AR-9202-28HRecalls & reported problems

Arthrex, Inc. · UNIVERS II REAMER, 8MM - HUDSON CONNECT · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867058460), received between May 22, 2024 and May 22, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Connection Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Connection Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 22, 2024Connection Problem; Connection Problem

    ON 4/5/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-9202-28H UNIVERS II REAMER HAD AN ISSUE. THIS CANAL REAMER "DEVICE" WAS ATTACHED TO A REAMER "T-HANDLE" IN A NORMAL MANNER AND WOULD NOT LOCK ONTO THE T HANDLE. A SECOND T HANDLE WAS USED AND IT WOULD NOT LOCK ONTO THE DEVICE AS WELL. THE NEXT SIZE LARGER CALAR REAMER WAS USED TO COMPLETE THE SURGICAL PROCEDURE IN A STANDARD FASHION. NO ADDED TIME OR ADVERSE EVENT OCCURRED. THIS OCCURRED DURING USE WITH NO PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED. ON 4/7/2023, THE SALES REPRESENTATIVE PROVIDED THE FOLLOWING INFORMATION VIA EMAIL: THIS OCCURRED DURING A REVERSE TSA PROCEDURE ON (B)(6) 2023. NO PIECE OF THE INSTRUMENT BROKE WHILE IT WAS NOT LOCKING. THE BONE QUALITY OF THE PATIENT WAS NOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19373898 ↗

Other INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices

All INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page