Skip to content

Arthrex® AR-300PD24Recalls & reported problems

Arthrex, Inc. · PIN DRIVER ATTACHMENT 1.0-2.4 MM · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00888867100176), received between Mar 28, 2024 and Sep 16, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Positioning Failure” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Positioning Failure2
  • Break1
  • Defective Device1
  • Detachment of Device or Device Component1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2026Positioning Failure; Use of Device Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50787474 ↗
  • MalfunctionApr 28, 2025Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT DURING A MICA SURGERY THE DEVICE DID NOT GRIP THE 2.4 K WIRES PROPERLY. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE FROM STRYKER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21919940 ↗
  • MalfunctionMay 23, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A MICA SURGERY THE ATTACHMENT DOES NOT HOLD THE WIRE FIRMLY AND SLIPS. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19388063 ↗
  • MalfunctionMay 23, 2024Break; Break

    IT WAS REPORTED THAT DURING A SURGERY A BOLT OF THE DEVICE FOR SECURING THE LOCK HAS FALLEN OUT. AS A RESULT, THE WIRE IS NO LONGER STUCK IN THE HOLDER. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE AVOE 28-APR-2023: IT WAS CONFIRMED THAT THE BOLT FELL OUT DURING A MICA (MINIMALLY INVASIVE CHEVRON AND AKIN) SURGERY. THE SURGERY WAS FINISHED SUCCESSFULLY BY INSERTING K-WIRES WITH A SMALLER DIAMETER AND CONTINUING THE DRILLING PROCESS. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19387830 ↗
  • MalfunctionMar 28, 2024Defective Device; Defective Device

    IT WAS REPORTED THAT THE LOCKING MECHANISM FOR THE WIRES IS TOO WEAK. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THERE WAS NO CASE INVOLVEMENT REPORTED. NO FURTHER INFORMATION RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18995122 ↗

Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices

All INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page