Skip to content

Arthrex® AR-513-BJ13Recalls & reported problems

Arthrex, Inc. · IBAL TKA,BEARING IMP,PS,SZ 8,13MM · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867107687), received between Feb 27, 2017 and Feb 27, 2017 — classified as 1 injury.

The most frequently reported specific device problem is “Loose or Intermittent Connection” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Loose or Intermittent Connection1

Patient problems

  • Ambulation or Postural Difficulties1
  • Edema1
  • Hematoma1
  • Limited Mobility Of The Implanted Joint1
  • No Code Available1
  • Pain1
  • Sleep Dysfunction1
  • Swelling1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 27, 2017Loose or Intermittent Connection

    IT WAS REPORTED THAT ON (B)(6) 2015 A PATIENT UNDERWENT A LEFT TKA PROCEDURE. ON (B)(6) 2017 THE SURGEON PERFORMED A REVISION LEFT TKA DUE TO UNEXPLAINED PAIN AND LOOSE TIBIA. DURING THE REVISION PROCEDURE THE SURGEON EXPLANTED THE (TIBIAL TRAY AR-503-TTTJ, LOT 1251021, ((B)(4)), BEARING IMPLANT AR-513-BJ13, LOT 113601404 ((B)(4)), FEMORAL IMPLANT AR-516-8L, LOT 108761335 ((B)(4)) AND PATELLA IMPLANT AR-504-PSC9, LOT 113601437 ((B)(4)). TO COMPLETE THE PROCEDURE THE SURGEON USED ANOTHER MANUFACTURER'S PRODUCT. PATIENT WAS A MALE, DOB (B)(6) 1956, WEIGHT (B)(6). PATIENT WAS IN THE OBESE RANGE BY BMI AT THE TIME OF INITIAL WHICH IS CONTRAINDICATED FOR TKA. PATIENT MEDICAL RECORDS DATED (B)(6) 2016 NOTED: LEFT KNEE X-RAY FINDINGS - PROSTHESIS IN PLACE IN PROPER ANATOMICAL ALIGNMENT WITHOUT ROTATION, LOOSENING OR STRESS SHIELDING. RECORDS NOTED PATIENT WOULD START WEIGHT LOSS PROGRAM TO SEE IF IT WOULD MAKE ANY DIFFERENCE. PATIENT MEDICAL RECORDS DATE (B)(6) 2016 NOTED: PATIENT HAD A I&D OF HEMATOMA (B)(6) 2015. PATIENT HAS HAD EMBOLIZATIONS X3, PATIENT WAS NOTED TO HAVE DECREASED MOBILITY, PAIN WITH SLEEPING, WALKING, STAIRS AND ALL WEIGHT BEARING ACTIVITIES. EXAMINATION OF THE LEFT K

    Patient problems: Ambulation or Postural Difficulties; Edema; Hematoma; Limited Mobility Of The Implanted Joint; No Code Available; Pain; Sleep Dysfunction; Swelling

    MAUDE report 6360430 ↗

Other PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLO/FEMORAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page