Skip to content

Arthrex® AR-600LRecalls & reported problems

Arthrex, Inc. · BATTERY PACK,UNSTRL,LI-ION FOR AR-600 · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
13
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00888867110946), received between Aug 31, 2023 and Jun 10, 2024 — classified as 13 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 5 reportsMay 2024: 4 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Split, Cut or Torn5
  • Break4
  • Material Rupture2
  • Crack1
  • Material Separation1
  • Overheating of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Insufficient Information4

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 10, 2024Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING A SURGERY THE BACK PART OF THE DEVICE BECAME HOT. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THIS WAS NOTICED BEFORE THE SURGERY. NO FURTHER INFORMATION RECEIVED. UPDATE AVOE 11-APR-2022: IT WAS CONFIRMED THAT THE ERROR OCCURRED ON THE 04-APR-2022 DURING A HIP SURGERY. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19499425 ↗
  • MalfunctionMay 9, 2024Break; Break

    ON 04/16/2024, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA (B)(4) THAT A QTY. 2 OF AR-600L BATTERY PACK HOUSINGS WERE COMPLETELY SPLIT AND APART. THIS WAS DISCOVERED WHILE TRYING TO REMOVE BATTERIES FROM THE BATTERY HOUSING DURING AN ACL REPAIR PROCEDURE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19280446 ↗
  • MalfunctionMay 9, 2024Break; Break

    ON 04/16/2024, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA SEMS (B)(4) THAT A QTY. 2 OF AR-600L BATTERY PACK HOUSINGS WERE COMPLETELY SPLIT AND APART. THIS WAS DISCOVERED WHILE TRYING TO REMOVE BATTERIES FROM THE BATTERY HOUSING DURING AN ACL REPAIR PROCEDURE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19275354 ↗
  • MalfunctionMay 8, 2024Break; Break

    ON (B)(6) 2024, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA SEMS (B)(4) THAT A QTY. 2 OF AR-600L BATTERY PACK HOUSINGS WERE COMPLETELY SPLIT AND APART. THIS WAS DISCOVERED WHILE TRYING TO REMOVE BATTERIES FROM THE BATTERY HOUSING DURING A DISTAL BICEPS REPAIR PROCEDURE ON (B)(6) 2024. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19274878 ↗
  • MalfunctionMay 8, 2024Break; Break

    ON 04/16/2024, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA SEMS-06541011 THAT A QTY. 2 OF AR-600L BATTERY PACK HOUSINGS WERE COMPLETELY SPLIT AND APART. THIS WAS DISCOVERED WHILE TRYING TO REMOVE BATTERIES FROM THE BATTERY HOUSING DURING A DISTAL BICEPS REPAIR PROCEDURE ON (B)(6) 2024. NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19274642 ↗
  • MalfunctionApr 19, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    ON 2/26/2024, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA (B)(4) THAT AN AR-600L BATTERY PACK HOUSING WAS COMPLETELY SPLIT AND APART. THIS WAS DISCOVERED AFTER A SHOULDER SCOPE WITH A LABRAL REPAIR PROCEDURE ON (B)(6) 2024.

    Patient problems: Insufficient Information

    MAUDE report 19144566 ↗
  • MalfunctionApr 19, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    ON (B)(6) 2024, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA (B)(4 )THAT AN AR-600L BATTERY PACK HOUSING WAS COMPLETELY SPLIT AND APART. THIS WAS DISCOVERED AFTER A FRACTURE PROCEDURE ON (B)(6) 2024.

    Patient problems: Insufficient Information

    MAUDE report 19143180 ↗
  • MalfunctionApr 19, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    ON 2/26/2024, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA (B)(4) THAT AN AR-600L BATTERY PACK HOUSING WAS COMPLETELY SPLIT AND APART. THIS WAS DISCOVERED AFTER A SHOULDER SCOPE WITH A LABRAL REPAIR PROCEDURE ON (B)(6) 2024.

    Patient problems: Insufficient Information

    MAUDE report 19142507 ↗
  • MalfunctionApr 19, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE HOUSING IS RUPTURED. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THERE WAS NO CASE INVOLVEMENT REPORTED. NO FURTHER INFORMATION RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19142245 ↗
  • MalfunctionApr 19, 2024Material Rupture; Material Rupture

    IT WAS REPORTED THAT THE HOUSING IS RUPTURED. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THERE WAS NO CASE INVOLVEMENT REPORTED. NO FURTHER INFORMATION RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19142188 ↗
  • MalfunctionMar 27, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    ON (B)(6) 2024, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA SEMS (B)(4) THAT AN AR-600L BATTERY PACK HOUSING WAS COMPLETELY SPLIT AND APART. THIS WAS DISCOVERED AFTER A SHOULDER SCOPE WITH A LABRAL REPAIR PROCEDURE ON (B)(6) 2024.

    Patient problems: Insufficient Information

    MAUDE report 18987884 ↗
  • MalfunctionOct 26, 2023Material Separation; Material Split, Cut or Torn; Material Separation; Material Split, Cut or Torn

    ON 9/29/2023, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA (B)(6) THAT AN AR-600L BATTERY PACK HOUSING WAS SPLIT ALONG THE SEAMS. THIS WAS DISCOVERED DURING THE BATTERY REMOVAL AFTER THE CASE, WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18011854 ↗
  • MalfunctionAug 31, 2023Crack; Crack

    ON (B)(6) 2023, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA SEMS THAT AN AR-600L BATTERY PACK HAS A CRACK ON THE HOUSING. IT WAS INVOLVED IN A CASE. PATIENT WAS NOT AFFECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17662314 ↗

Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices

All INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices →

Sources and updates