Arthrex® AR-8332HRecalls & reported problems
Arthrex, Inc. · SHAVER HP, SYNERGY, HANDCONTROL · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Arthrex Synergy shaver handpiece is a powered surgical instrument used to shave bone and tissue during orthopedic procedures. The 65 reports mostly describe the handpiece overheating, cords and wiring splitting or being cut, and positioning failures. There are no deaths and no injuries. No recalls name this device. Reports of these malfunctions rose from 2 in the earlier 12-month period to 17 in the most recent 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 62
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 62 adverse event reports that list this exact device (DI 00888867123991), received between Mar 14, 2022 and Sep 16, 2026 — classified as 62 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (16 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device16
- Material Split, Cut or Torn8
- Use of Device Problem5
- Defective Device4
- Excessive Heating4
- Positioning Failure4
- Physical Resistance/Sticking3
- Difficult to Remove2
Patient problems
- No Clinical Signs, Symptoms or Conditions56
- Insufficient Information7
- Laceration(s)1
Recent adverse event reports
The 25 most recent of 62 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2026Material Frayed
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50788198 ↗MalfunctionApr 16, 2026Overheating of Device
ON 27-MAR-2026, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT THE AR-8332H SHAVER HANDPIECE IS OVERHEATING. THIS ISSUE WAS DISCOVERED WITH NO CASE INVOLVEMENT.
Patient problems: Insufficient Information
MAUDE report 24896493 ↗MalfunctionApr 16, 2026Material Split, Cut or Torn
ON 27-MAR-2026, A FACILITIES REPRESENTATIVE REPORTED VIA (B)(4) THAT THE AR-8332H SHAVER HANDPIECE HAS A CUT IN THE CORD. THIS ISSUE WAS DISCOVERED AFTER THE CASE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24895493 ↗MalfunctionApr 9, 2026Material Separation
ON 19-MAR-2026, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-8332H SHAVER HAD EXPOSED WIRES AT THE BASE. A CASE WAS INVOLVED WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24833076 ↗MalfunctionApr 9, 2026Material Split, Cut or Torn
ON 18-MAR-2026, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-8332H SHAVER HANDCONTROL HAD A CUT IN THE CORD. THIS WAS DISCOVERED DURING A CASE WITH NO ADVERSE EFFECTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24832735 ↗MalfunctionMar 16, 2026Material Split, Cut or Torn
ON (B)(6), A SALES REPRESENTATIVE REPORTED VIA SEMS-(B)(4) THAT THE AR-8332H SHAVER HANDPIECE HAS A SLICE IN THE CORD. THIS ISSUE WAS DISCOVERED DURING A CASE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24606044 ↗MalfunctionMar 5, 2026Positioning Failure; Use of Device Problem
ON 12-FEB-2026, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT THE AR-8332H SHAVER HANDPIECE IS HAVING ISSUES PLACING A BLADE INSIDE. THIS ISSUE WAS DISCOVERED DURING A CASE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24519155 ↗MalfunctionFeb 26, 2026Positioning Failure; Use of Device Problem
ON 04-FEB-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-8332H SHAVER IS NOT SNAPPING INTO PLACE PROPERLY. DISCOVERED DURING A CASE WITH NO PATIENT EFFECT. ADDITIONAL INFO PROVIDED 25-FEB-26: ISSUE OCCURRED IN THE SURGICAL FIELD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24454790 ↗MalfunctionFeb 26, 2026Positioning Failure
ON 04-FEB-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-8332H SHAVER IS NOT SNAPPING INTO PLACE PROPERLY. DISCOVERED DURING A CASE WITH NO PATIENT EFFECT. ADDITIONAL INFO PROVIDED 25-FEB-26: ISSUE OCCURRED IN THE SURGICAL FIELD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24454706 ↗MalfunctionJan 19, 2026Electrical /Electronic Property Problem
ON 31-DEC-2025, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA (B)(4) THAT AN AR-8332H SHAVER HANDPIECE HAD AN ELECTRICAL ISSUE. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24110517 ↗MalfunctionDec 29, 2025Use of Device Problem; Connection Problem; Use of Device Problem; Connection Problem
IT WAS REPORTED THAT DURING AN ACROMIOCLAVICULA (AC) DECOMPRESSION SURGERY THE SHAVER HANDPIECE AR-8332H (SN (B)(6)) DAMAGED THE BURR AR-8400OBE (LOT: 15496415). THE BURR GOT TORN UP IN THE BOTTOM WHERE IT WAS CONNECT TO THE SHAVER. PLASTIC PIECES REMAINED IN THE SHAFT OF THE SHAVER HANDPIECE. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23907105 ↗MalfunctionDec 29, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT DURING AN ACROMIOCLAVICULA (AC) RESECTION SURGERY THE SHAVER HANDPIECE AR-8332H (SN (B)(6) DAMAGED THE BURR AR-8400OBE (LOT: 15446780). THE BURR GOT TORN UP IN THE BOTTOM WHERE IT WAS CONNECTED TO THE SHAVER. PLASTIC PIECES REMAINED IN THE SHAFT OF THE SHAVER HANDPIECE. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23904997 ↗MalfunctionNov 21, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
ON 10/30/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-8332H SHAVER HAS AN UNKNOWN OBJECT STUCK INSIDE THE HANDPIECE. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23619777 ↗MalfunctionNov 21, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT DURING A SHOULDER DECOMPRESSION SURGERY THE SHAVER HANDPIECE AR-8332H (SN: (B)(6)) DAMAGED 2 BURRS AR-8400OBE (BATCH 15446780). THEY GOT TORN UP IN THE BOTTOM WHERE THEY ATTACH TO SHAVER HANDPIECE. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23619003 ↗MalfunctionMay 13, 2025Material Split, Cut or Torn; Material Split, Cut or Torn
ON 04/24/2025, A FACILITY REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-8332H SYNERGY SHAVER CORD WAS CUT. THE CASE INVOLVEMENT IS UNKNOWN.
Patient problems: Insufficient Information
MAUDE report 22012445 ↗MalfunctionFeb 12, 2025Unintended Movement; Unintended Movement
ON 01/21/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-8332H SHAVER HANDPIECE WOBBLED WHEN IN USE. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21366775 ↗MalfunctionSep 17, 2024Positioning Failure; Positioning Failure
IT WAS REPORTED THAT DURING A KNEE JOINT ARTHROSCOPY THE DEVICE WAS DIFFICULT TO LOCK AND DID NOT GIVE THE SAME PERFORMANCE DURING SURGERY. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20238616 ↗MalfunctionSep 10, 2024Material Deformation; Material Deformation
ON (B)(6)2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS-06641198 THAT AN AR-8332H SHAVER HAS EXPOSED WIRING ON THE CABLE. DISCOVERED DURING A CASE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20181646 ↗MalfunctionJul 18, 2024Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED THAT THE SHAVER HAS A LOOSE CONTACT. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THERE WAS NO CASE INVOLVEMENT REPORTED. NO FURTHER INFORMATION RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19773078 ↗MalfunctionJun 24, 2024Device Contamination with Body Fluid; Device Contamination with Body Fluid
ON 05/28/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-8332H SHAVER WAS FOUND WITH BIO-BURDEN IN THEM. DISCOVERED IN CASE, PATIENT NOT AFFECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19593752 ↗MalfunctionJun 21, 2024Environmental Particulates; Environmental Particulates
ON 05/28/2024 IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-8332H SHAVER WAS FOUND WITH BIO-BURDEN IN THEM. DISCOVERED IN CASE, PATIENT NOT AFFECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19588856 ↗MalfunctionJun 19, 2024Material Deformation; Material Deformation
ON 05/23/2024, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA (B)(4) THAT AN AR-8332H SHAVER HP, HAS EXPOSED WIRES. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19567968 ↗MalfunctionJun 14, 2024Excessive Heating; Excessive Heating
IT WAS REPORTED THAT THE DEVICE GETS HOT. NO ADDITIONAL INFORMATION REGARDING A SPECIFIC FAILURE MODE WAS PROVIDED. THERE WAS NO HARM FOR THE PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19535505 ↗MalfunctionJun 7, 2024Device-Device Incompatibility; Device-Device Incompatibility
ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-8332H SHAVER HP WAS CAUSING INTERFERENCE. IT INTERFERED WITH THE PATIENT'S EKG MONITOR WHEN USED IN THE CASE. THIS WAS DISCOVERED DURING AN UNSPECIFIED PROCEDURE, WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19489426 ↗MalfunctionMay 23, 2024Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE DEVICE GETS HOT. THERE WAS NO HARM FOR THE PATIENT, OPERATOR OR THIRD PARTY REPORTED. NO FURTHER INFORMATION RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19388688 ↗
FDA clearance and classification
The FDA regulates product code HWE (instrument, surgical, orthopedic, ac-powered motor and accessory/attachment) under 21 CFR 878.4820 as a Class I device. 21 CFR 878.4820 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Arthrex® AR-8332H
Has the Arthrex® AR-8332H been recalled?
No FDA recall record in our data names this exact device identifier (DI 00888867123991), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Arthrex® AR-8332H?
The FDA's MAUDE database lists 62 adverse event reports for this device, received between Mar 14, 2022 and Sep 16, 2026. In those reports, the most common specific device problems are “Overheating of Device” (16), “Material Split, Cut or Torn” (8) and “Use of Device Problem” (5); the most common patient problem is “Laceration(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Arthrex® AR-8332H?
None of the reports are classified as deaths, 0 as injuries and 62 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Arthrex® AR-8332H FDA approved?
Devices of this type (product code HWE) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Arthrex, Inc.
Who makes the Arthrex® AR-8332H, and is it still sold?
The FDA's device database lists Arthrex, Inc. as the company responsible for it (brand Arthrex®, model AR-8332H). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices
- NA 511.701Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT2,160 reports
- SmartLife 7126-120-000Stryker Corporation · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT1,065 reports
- NA 05.001.224Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT970 reports
- NA 05.001.206Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT622 reports
- NA 532.021Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT518 reports
- NA 2570-04-501Depuy (Ireland) · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT500 reports
- NA 532.110Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT432 reports
- NA 05.001.223Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT393 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page