Arthrex® AR-8948SRecalls & reported problems
Arthrex, Inc. · BioSync Wedge Instrumentation Set · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867217478), received between Jun 12, 2024 and Jun 18, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Material Separation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation1
- Pitted1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 18, 2024Pitted; Pitted
IT WAS REPORTED, THE SALES REP, NOTICED THAT ALL THE HANDLES UNSCREW FROM THE TRAIL WEDGE PIECES. THE SALES REP BROUGHT THE TRAY BACK TO THE DISTRIBUTORS OFFICE AND MANY OF THE THREADS HAVE WHAT APPEARED TO BE ALMOST CORROSIVE CHARACTERISTICS EMBEDDED IN THE THREADS AND IT APPEARED AS IF THESE TRIALS HADN¿T BEEN REMOVED FROM THEIR RESPECTIVE HANDLES IN YEARS.
Patient problems: Insufficient Information
MAUDE report 19564194 ↗MalfunctionJun 12, 2024Material Separation; Material Separation
IT WAS REPORTED THAT DURING AN EVANS OSTEOTOMY, THE TRIALS FROM THE BIOSYNC TRAY UN-THREADED FROM THEIR INSERTER HANDLE. TWO DIFFERENT TRIALS WERE USED AT ONE POINT IN THE CASE WHILE TRIALING FOR THE ALLOSYNC WEDGE SIZE THE SURGEON WANTED TO IMPLANT. THE CASE WAS COMPLETED AND WAS NOT AFFECTED IN ANY WAY NEGATIVELY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19519920 ↗
Other INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices
- Arthrex® AR-13995NArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE430 reports
- Arthrex® AR-12990NArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE197 reports
- Arthrex® AR-13991NArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE128 reports
- Arthrex® AR-8981-08SArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall3 reports
- Arthrex® AR-8981-06SArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall1 report
- Centurion EB16Centurion Medical Products Corporation · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall0 reports
- NA Lighthandle CoverSteris Corporation · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall0 reports
- Centurion Medical Products Corp SUT15830AMedline Industries, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page