Arthrex® AR-521-TBA8Recalls & reported problems

Arthrex, Inc. · UKA TIBIAL BEARING, SIZE 1, 8MM, VIT-E · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00888867231283), received between Dec 16, 2019 and Jul 10, 2025 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Fitting Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fitting Problem3
  • Positioning Failure2
  • Failure to Advance1
  • Naturally Worn1
  • Unstable1

Patient problems

  • Insufficient Information4
  • Pain1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 10, 2025Fitting Problem; Fitting Problem

    ON (B)(6) 2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT TWO AR-521-TBA8 IBALANCE UKA TIBIAL BEARING IMPLANTS FAILED TO SNAP INTO THE TRAY. MULTIPLE TECHNIQUES WERE ATTEMPTED, AND THE TRAY WAS CONFIRMED TO BE FREE OF DEBRIS; HOWEVER, NEITHER 8MM TIBIAL BEARINGS COULD BE SECURED INTO THE LOCKING TABS. THE COMPONENT REMAINED PROUD EITHER AT THE CENTRAL PORTION OR ON THE MEDIAL SIDE OF THE TRAY. THIS ISSUE WAS IDENTIFIED DURING A RIGHT MEDIAL UKA PROCEDURE ON (B)(6) 2025. ADDITIONAL INFORMATION REQUESTED ON 6/25/2025.

    Patient problems: Insufficient Information

    MAUDE report 22469173 ↗
  • MalfunctionApr 25, 2024Failure to Advance; Failure to Advance

    ON (B)(6) 2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-521-TBA8 UKA TIBIAL BEARING IMPLANT WOULD NOT GO IN. SURGEON OPENED A NEW PRODUCT AND CASE WAS COMPLETED. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2022.

    Patient problems: Insufficient Information

    MAUDE report 19179543 ↗
  • MalfunctionMar 28, 2024Positioning Failure; Fitting Problem; Positioning Failure; Fitting Problem

    ON 03/13/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA E-MAIL THAT ((2) AR-521-TBA8 IBALANCE® UKA TIBIAL BEARING IMPLANTS WERE NOT LOCKING INTO THE AR-511-T1L IBALANCE® UKA TIBIAL TRAY. THIS OCCURRED DURING A CASE. NO ADDITIONAL INFORMATION WAS PROVIDED, AND ADDITIONAL INFORMATION HAS BEEN ASKED.

    Patient problems: Insufficient Information

    MAUDE report 18997644 ↗
  • MalfunctionMar 28, 2024Positioning Failure; Fitting Problem; Positioning Failure; Fitting Problem

    ON 03/13/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA E-MAIL THAT ((2) AR-521-TBA8 IBALANCE® UKA TIBIAL BEARING IMPLANTS WERE NOT LOCKING INTO THE AR-511-T1L IBALANCE® UKA TIBIAL TRAY. THIS OCCURRED DURING A CASE. NO ADDITIONAL INFORMATION WAS PROVIDED, AND ADDITIONAL INFORMATION HAS BEEN ASKED.

    Patient problems: Insufficient Information

    MAUDE report 18994410 ↗
  • InjuryDec 16, 2019Unstable; Naturally Worn

    THE PATIENT HAD ORIGINAL SURGERY FOR A UNI KNEE REPLACEMENT ON (B)(6) 2018. SHE REPORTED PAIN AND INSTABILITY; XRAYS SHOWED THAT THE POLY HAD WORN DOWN. ARTHREX PARTS AR-521-TBA8, AR-501-UFLD AND AR-501-TTLA WERE REMOVED. A TOTAL KNEE REVISION WAS DONE ON (B)(6) 2019 WITH A COMPETITOR'S DEVICE. PATIENT FEMALE, 60S.

    Patient problems: Pain

    MAUDE report 9473603 ↗

Other PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices →