Arthrex® AR-1610LGRecalls & reported problems
Arthrex, Inc. · LOCKING GUIDE · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867241947), received between Dec 22, 2023 and Feb 8, 2024 — classified as 2 injuries.
The most frequently reported specific device problem is “Incorrect Measurement” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect Measurement2
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Bone Fracture(s)1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryFeb 8, 2024Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT DURING A MENISCAL ROOT REPAIR SURGERY ON TWO SEPARATE ATTEMPTS THE DRILL CAME OUT 5MM MEDIALLY FROM THE POINT AT WHICH IT SHOULD HAVE COME OUT AS INDICATED BY THE 7.5MM OFFSET MARK ON THE HOOK. THIS LED TO AN UNNECESSARY LOSS OF TIBIA BONE IN THE PATIENT. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE (AR-1610LG). IT WAS NECESSARY TO USE THE TIBIA ACL (HALO) JIG TO ACCURATELY DRILL THE TIBIA TUNNEL. UPDATE (B)(4) 11-JAN-2024: FURTHER INFORMATION REVEALED THAT AR-1610LG BATCH: 43212209 WAS ALSO INVOLVED IN THIS INCIDENT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 18666667 ↗InjuryDec 22, 2023Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT DURING A MENISCAL ROOT REPAIR SURGERY ON TWO SEPARATE ATTEMPTS THE DRILL CAME OUT 5MM MEDIALLY FROM THE POINT AT WHICH IT SHOULD HAVE COME OUT AS INDICATED BY THE 7.5MM OFFSET MARK ON THE HOOK. THIS LED TO AN UNNECESSARY LOSS OF TIBIA BONE IN THE PATIENT. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A NEW DEVICE WITH THE SAME PART NUMBER. IT WAS NECESSARY TO USE THE TIBIA ACL (HALO) JIG TO ACCURATELY DRILL THE TIBIA TUNNEL.
Patient problems: Bone Fracture(s)
MAUDE report 18387376 ↗
Other INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices
- Arthrex® AR-13995NArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE430 reports
- Arthrex® AR-12990NArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE197 reports
- Arthrex® AR-13991NArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE128 reports
- Arthrex® AR-8981-08SArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall3 reports
- Arthrex® AR-8981-06SArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall1 report
- HarmonyAIR LAS0042Steris Corporation · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall0 reports
- Centurion EB16Centurion Medical Products Corporation · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall0 reports
- Centurion Medical Products Corp SUT15830AMedline Industries, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page