Skip to content

Arthrex® AR-200SPRecalls & reported problems

Arthrex, Inc. · SPRAYCLIP FOR AR-200 · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867245822), received between Oct 8, 2024 and Jan 8, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Overheating of Device1

Patient problems

  • Full thickness (Third Degree) Burn1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 8, 2026Detachment of Device or Device Component

    ON 16-DEC-2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT THE METAL SPRAY COMPONENT OF THE AR-200SP CAME OFF AND WOULD NOT REATTACH. NOTICED DURING A CASE, DEVICE SWAPPED, AND CASE COMPLETED WITH NO PATIENT EFFECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24001864 ↗
  • InjuryOct 8, 2024Overheating of Device; Overheating of Device

    ON 10TH SEPTEMBER 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-200M MOTOR WITH CABLE 3.5M WAS HEATING UP. THIS WAS DETECTED DURING A TOTAL ANKLE REPLACEMENT PROCEDURE ON (B)(6) 2024. DR. (B)(6) WAS DOING A TOTAL ANKLE REPLACEMENT +/- HEEL SHIFT. ARTHREX MIS HANDPIECE WAS ON THE SETUP FOR THE +/- HEEL SHIFT PART OF THE PROCEDURE. AFTER THE CASE, THE SALES REPRESENTATIVE WAS TOLD BY THE NURSING STAFF THAT THE ARTHREX MIS HANDPIECE HAD CAUSED SEVERE BURNS TO THE PATIENT'S THIGH AREA. WHEN THE SALES REPRESENTATIVE WENT TO THE THEATRE, HE WAS TOLD BY THE SCOUT NURSE THAT THE SCOUT NURSE "ACCIDENTALLY PLUGGED THE ARTHREX MIS HANDPIECE INTO THE SYNTHES E-PEN CONSOLE BECAUSE IT FIT IN IT PERFECTLY". THAT CAUSED THE HANDPIECE TO HEAT UP AND MELT THROUGH THE HOLDING QUIVER, CAUSING THE BURNS TO THE PATIENT. SALES REPRESENTATIVE WAS TOLD BY THE SCRUB NURSE THAT NO ONE NOTICED IT HEATING UP BECAUSE THE HANDPIECE WAS NOT USED DURING THE CASE AS HEEL SHIFT WAS NOT REQUIRED AT THE END. SALES REPRESENTATIVE'S ATTENDANCE WAS NOT REQUIRED FOR THIS CASE AS BOTH NURSING STAFF ARE TRAINED AND KNOW HOW TO PLUG IN ARTHREX MIS HANDPIECE. UPDATE ON PATIENT CONDITION AS OF (B)(6) 202

    Patient problems: Full thickness (Third Degree) Burn

    MAUDE report 20396664 ↗

Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices

All INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices →