The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00888867259782), received between Aug 18, 2022 and Jan 9, 2025 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Break” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break8
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Insufficient Information4
  • Foreign Body In Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 9, 2025Break; Break

    IT WAS REPORTED THAT DURING A CUFF SURGERY THE TIP BROKE. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH ANOTHER DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21114641 ↗
  • MalfunctionJun 7, 2024Break; Break

    ON 12/15/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-8500RFOE RETRACTABLE FLUSHCUT BURR HAS BROKEN OFF IN THE PATIENT'S SHOULDER. SURGEON WAS ABLE TO REMOVE THE BROKEN TIP. THIS WAS DISCOVERED DURING A PROCEDURE. ADDITIONAL INFORMATION RECEIVED ON 12/16/2022: THIS WAS DISCOVERED DURING A SHOULDER ARTHROSCOPY PROCEDURE ON (B)(6) 2022. PER SALES REPRESENTATIVE, TWO AR-8500RFOE RETRACTABLE FLUSHCUT BURR BROKE CONSECUTIVELY DURING THE PROCEDURE. ALL BROKEN FRAGMENTS FROM BOTH BURR'S WERE RETRIEVED FROM INSIDE THE PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 19482650 ↗
  • MalfunctionDec 22, 2023Break; Break

    IT WAS REPORTED THAT DURING A SHOULDER SURGERY THE TIP OF THE DEVICE BROKE OFF WHILE MOVING THE TIP FORWARD. THE BROKEN TIP WAS RETRIEVED OUT OF THE PATIENT. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18385391 ↗
  • MalfunctionDec 22, 2023Break; Break

    IT WAS REPORTED THAT DURING A SHOULDER SURGERY THE TIP OF THE DEVICE BROKE OFF WHILE MOVING THE TIP FORWARD. THE BROKEN TIP WAS RETRIEVED OUT OF THE PATIENT. THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT DEVICE. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18385356 ↗
  • MalfunctionApr 28, 2023Break; Break

    IT WAS REPORTED, THAT DURING A ROTATOR CUFF SURGERY, THE BURR TIP BROKE OFF. THE BROKEN PART OF THE DEVICE WAS REMOVED FROM THE PATIENT WITH AN INSTRUMENT RETRIEVER. ACCORDING TO THE SURGEON NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY OCCURRED. THE SURGERY WAS FINISHED SUCCESSFULLY WITH ANOTHER SHAVER. IT WAS NOT NECESSARY TO SWITCH THE SURGICAL TECHNIQUE OR DO A SECOND SURGERY. UPDATE, KHE 19-APR-2023, 2 ORIGINAL PACKED DEVICES WERE RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16837808 ↗
  • MalfunctionJan 3, 2023Break

    ON 12/15/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-8500RFOE RETRACTABLE FLUSHCUT BURR HAS BROKEN OFF IN THE PATIENT'S SHOULDER. SURGEON WAS ABLE TO REMOVE THE BROKEN TIP. THIS WAS DISCOVERED DURING A PROCEDURE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16086990 ↗
  • MalfunctionAug 29, 2022Break

    IT WAS REPORTED ON (B)(6) 2022 BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-8500RFOE BURR BROKE OFF THE SHAFT. THIS WAS DISCOVERED DURING A CASE AND BROKE INSIDE THE JOINT SPACE.

    Patient problems: Foreign Body In Patient

    MAUDE report 15310783 ↗
  • MalfunctionAug 29, 2022Break

    IT WAS REPORTED ON 08/11/2022 BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-8500RFOE BURR BROKE OFF THE SHAFT. THIS WAS DISCOVERED DURING A CASE AND BROKE INSIDE THE JOINT SPACE.

    Patient problems: Insufficient Information

    MAUDE report 15310747 ↗
  • MalfunctionAug 18, 2022Detachment of Device or Device Component

    ON (B)(6) 2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-8500RF0E RETRACTABLE FLUSHCUT BURR HEAD CAME OFF OF THE JOINT AND THE SURGEON HAD TO SEARCH FOR IT. THIS WAS DISCOVERED DURING AN UNSPECIFIED PROCEDURE. ADDITIONAL INFORMATION REQUESTED.

    Patient problems: Insufficient Information