Arthrex® AR-8750-01Recalls & reported problems
Arthrex, Inc. · DPTH DEV, CANN SCRWS · Class 1
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867263604), received between Dec 11, 2025 and Dec 11, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Failure to Clean Adequately” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Clean Adequately1
Patient problems
- Bacterial Infection1
- Insufficient Information1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryDec 11, 2025Failure to Clean Adequately; Failure to Clean Adequately
ON 5TH DEC 2025, IT WAS REPORTED BY AN ARTHREX'S EMPLOYEE VIA AN EMAIL THAT FOR 1 UNIT OF AR-8750-01, DEPTH DEVICE FOR CANNULATED SCREWS, BLOOD CLOT WAS FOUND IN THE INSTRUMENT DEPTH GAUGE. THIS WAS DETECTED DURING A SPRING LIGAMENT RECON PROCEDURE ON (B)(6) 2025. THE SURGEON CONTINUED USING THE AR-8750-01 AFTER SOAKING IT IN CHLORHEX. ADDITIONAL INFORMATION: THE HOSPITAL INFORMED ARTHREX SINGAPORE ON (B)(6) 2025 THAT THE PATIENT SUFFERED FROM SIDE EFFECT AFTER THE SURGERY. ADDITIONAL INFORMATION RECEIVED ON 10TH DEC 2025: THE PATIENT WENT BACK TO THE HOSPITAL FOR WOUND WASHOUT 3 TIMES DUE TO INFECTION.
Patient problems: Bacterial Infection; Insufficient Information
MAUDE report 23779822 ↗
Other GAUGE, DEPTH devices
- NA 319.006Synthes GmbH · GAUGE, DEPTH899 reports
- NA 03.133.080Synthes GmbH · GAUGE, DEPTH174 reports
- NA 03.010.072Synthes GmbH · GAUGE, DEPTH133 reports
- MULTILOC 03.019.017Synthes GmbH · GAUGE, DEPTH119 reports
- NA 319.01Synthes GmbH · GAUGE, DEPTH113 reports
- NA 319.004Synthes GmbH · GAUGE, DEPTH82 reports
- QUICKSET 2274-60-000Depuy (Ireland) · GAUGE, DEPTH68 reports
- NA 319.09Synthes GmbH · GAUGE, DEPTH65 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.