Arthrex® AR-8990DSRecalls & reported problems

Arthrex, Inc. · Disposables Kit, for DX FiberTak · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00888867264984), received between Sep 19, 2022 and Jul 30, 2026 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Break” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break7
  • Material Twisted/Bent3
  • Dull, Blunt1

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • Insufficient Information2

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 30, 2026Material Twisted/Bent

    Patient problems: Insufficient Information

    MAUDE report 50085247 ↗
  • MalfunctionMay 15, 2026Break

    ON 24-APR-2026, AN ARTHREX SUBSIDIARY EMPLOYEE RETURNED AN AR-8990ST FIBERTAK DX SUTURE ANCHOR THAT BROKE DURING INSERTION INTO THE BONE. ADDITIONALLY, AR-8990DS DISPOSABLE KIT FOR DX FIBERTAK RETURNED BECAUSE THE 2 COMPONENTS WERE STUCK TOGETHER AND COULD NOT BE DISASSEMBLED. THE TIP OF THE DEVICE'S GUIDEWIRE WAS BENT. THERE WAS NO SIGNIFICANT CASE DELAY, NO ADDITIONAL ANESTHESIA ADMINISTERED, AND NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25187343 ↗
  • MalfunctionNov 4, 2025Break; Break

    ON 10/27/2025, A FACILITY REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-8990DS FIBERTAK DX SUTURE ANCHOR DISPOSABLES KIT DRILL BIT TIP BROKE WHEN DRILLING DURING A DELTOID REPAIR. THE CASE WAS COMPLETED BY OPENING ANOTHER DRILL BIT. THE BROKEN PIECE WAS RETRIEVED FROM THE PATIENT WITHOUT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23466754 ↗
  • MalfunctionFeb 24, 2025Break; Break

    ON 02/11/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-8990DS DISPOSABLES KIT BROKE. THIS OCCURRED DURING A CASE WHILE DRILLING. THERE WAS NO ADDITIONAL INFORMATION PROVIDED, AND ADDITIONAL INFORMATION HAS BEEN REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21446595 ↗
  • MalfunctionNov 8, 2024Dull, Blunt; Dull, Blunt

    ON 10/24/2024, A SALES REPRESENTATIVE REPORTED VIA (B)(6) THAT AN AR-8990DS DISPOSABLES KIT FOR FIBERTAK HAD DULL DRILL BITS. ANOTHER DISPOSABLE KIT WAS OPENED TO COMPLETE THE CASE, WITH NO ADVERSE EFFECTS ON THE PATIENT. THIS WAS DISCOVERED DURING AN ANKLE DELTOID REPAIR PROCEDURE ON (B)(6) 2024, WITH NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20641511 ↗
  • MalfunctionSep 6, 2024Break; Break

    ON 08/15/2024, A DISTRIBUTOR REPORTED VIA (B)(4) THAT AN AR-8990DS DISPOSABLES KIT FOR FIBERTAK DX SUTURE ANCHOR HAD THE 1.6 DRILL BROKEN, WITH NO EFFECT ON THE PATIENT. THIS WAS DISCOVERED DURING A PROCEDURE, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION WAS REQUESTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20162534 ↗
  • MalfunctionJun 7, 2024Break; Break

    ON 10/1/2021, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT (2) AR-8990ST FIBERTAK SUTURE ANCHORS WOULD NOT DEPLOY AND AN AR-8990DS DRILL GUIDE AND GUIDE WIRE KIT WAS DAMAGED. SURGEON HAS TO USE LARGER ANCHORS TO COMPLETE CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19490331 ↗
  • MalfunctionMay 21, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED DURING SURGERY THE K-1 WIRE DRILL WITHIN THE AR-8990DS BENT. THE PROCEDURE WAS COMPLETED WITHOUT ISSUE AFTER RE-DRILLING WITH NEW KIT AND ANCHOR. THE REP REPORTED THE DEVICES DID NOT BREAK INTO DISCRETE PIECES. THE SUTURE ON THE ANCHOR UNLOADED AND WAS PIERCED BY THE FORKED TIP OF THE ANCHOR RENDERING IT UNUSABLE.

    Patient problems: Insufficient Information

    MAUDE report 19357748 ↗
  • MalfunctionApr 30, 2024Material Twisted/Bent; Material Twisted/Bent

    ON 09/22/2021 IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA (B)(4) THAT AN AR-8990DS DISPOSABLES KIT WHEN OPENED AND ATTEMPTED TO ATTACH A 1.6 MM DRILL, IT WAS BENT. THIS WAS DISCOVERED DURING A PROCEDURE, THE PROCEDURE WAS COMPLETED USING A NEW PRODUCT, NO PATIENT EFFECT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19215050 ↗
  • MalfunctionSep 29, 2023Break; Break

    ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS-06019604 THAT AN AR-8990DS DISPOSABLES KIT, FOR DX FIBERTAK HAD A DRILL BIT BROKE OFF IN THE PATIENT'S PATELLA. ALL FRAGMENTS WERE NOT RETRIEVED FROM THE PATIENT WITH NO FURTHER DELAY. THE CASE WAS COMPLETED WITH NO ISSUES AND NO FURTHER INFORMATION PROVIDED. THIS WAS DISCOVERED DURING A MPFL PROCEDURE ON (B)(6) 2023, WITH NO REPORTED PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17842899 ↗
  • MalfunctionSep 19, 2022Break

    IT WAS REPORTED ON (B)(6) 2022 BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-8990DS DISPOSABLES KIT WIRE BROKE WHILE DRILLING. THIS WAS DISCOVERED DURING A CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15445611 ↗

Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices

All INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices →