Skip to content

Arthrex® AR-1938PKRecalls & reported problems

Arthrex, Inc. · PERC INSERT KIT FOR 3.0 MM KNOTLESS ST · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867271647), received between Feb 13, 2020 and Jun 18, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Material Twisted/Bent1
  • Physical Resistance/Sticking1

Patient problems

  • Device Embedded In Tissue or Plaque1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 18, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED ON (B)(6) 2023 BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-1938PK PERC INSERT KIT HAD A DRILL BIT THAT WAS JAMMED IN THE DRILL SLEEVE. THIS OCCURRED ON (B)(6) 2023 AT AN UNSPECIFIED TIME WITHOUT ANY CASE INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19558633 ↗
  • InjuryFeb 13, 2020Break; Material Twisted/Bent

    IT WAS REPORTED THAT THE NITINOL WIRE BENT AND BROKE WHEN THE GUIDE WAS REMOVED FROM THE PATIENT. A PARTIAL WIRE, ABOUT 2" WAS LEFT IN THE PATIENT. THE PATIENT IS TO RETURN ON (B)(6) 2020 TO HAVE IT REMOVED. WENT TO TAKE THE GUIDE OUT AND IT BENT. THERE WAS NO REP IN THE CASE, PATIENT IS MALE; LABRUM CASE. ADDITIONAL INFORMATION OBTAINED 2/13/20: THE WIRE FRAGMENT WAS SUCCESSFULLY REMOVED ON (B)(6) 2020 WITH NO COMPLICATIONS.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 9706017 ↗