Skip to content

Arthrex® AR-11600NRFRecalls & reported problems

Arthrex, Inc. · Grspr, Algr, 2.75, NR w/FlushPort · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867273047), received between Oct 24, 2024 and Oct 24, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 24, 2024Break; Break

    IT WAS REPORTED THAT THE DEVICE IS DEFECTIVE. ALSO, THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY REPORTED. IT WILL BE PROCESSED AS REPAIR, NOT AS COMPLAINT. UPDATE (B)(6) 18-OCT-2024: FURTHER INFORMATION WAS PROVIDED THAT DURING THE REPORTED EVENT HAS OCCURRED DURING A SURGERY. THE SURGEON TRIED TO REMOVE A FREE JOINT BODY FROM THE JOINT. THE FORCEPS BROKE OFF DURING THE FIRST ATTEMPT. UPDATE (B)(6) 24-OCT-2024: FURTHER INFORMATION WAS PROVIDED THAT THE SHEAR PIN INSIDE THE HANDLE BROKE OUTSIDE THE PATIENT. THE REPORTED EVENT OCCURRED DURING A KNEE ARTHROSCOPY WITH ANTERIOR CRUCIATE LIGAMENT PLASTIC. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. THE SURGERY WAS FINISHED SUCCESSFULLY WITH A DIFFERENT PLIER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20522669 ↗