Arthrex® AR-5815Recalls & reported problems
Arthrex, Inc. · Suture Tensioner/Cutter for knees · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00888867275676), received between Feb 11, 2020 and Oct 20, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Material Split, Cut or Torn1
- Use of Device Problem1
Patient problems
- Foreign Body In Patient1
- Insufficient Information1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 20, 2025Break; Use of Device Problem; Break; Use of Device Problem
ON 7TH OCTOBER 2025, IT WAS REPORTED BY AN ARTHREX EMPLOYEE VIA EMAIL THAT FOR AN AR-5815, KNOT PUSHER / SUTURE CUTTER, THE NITINOL FLAG SNAPPED SO THE SUTURE COULD NOT BE LOADED. THIS WAS DETECTED DURING USE IN A KNEE MENISCAL REPAIR PROCEDURE ON (B)(6) 2025. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AS AN ADDITIONAL SUTURE CUTTER WAS USED.
Patient problems: Insufficient Information
MAUDE report 23335205 ↗InjuryFeb 11, 2020Material Split, Cut or Torn
IT WAS REPORTED THAT DURING A KNEE ARTHROSCOPY PROCEDURE, THE SURGEON IMPLANTED A MENISCAL CINCH II PER THE TECHNIQUE. WHEN PUSHING THE KNOTS WITH THE SUTURE TENSIONER/ CUTTER FOR 2-0 FW, AR-5815, THE SUTURE WAS CUT. THE SURGEON REMOVED THE CUT SUTURE FROM THE JOINT WITH FORCEPS. THIS ADDED A 30 MIN DELAY TO THE CASE. THE CASE WAS COMPLETED WITH NO ADVERSE EFFECT TO THE PATIENT. ADDITIONAL INFORMATION REQUESTED. ADDITIONAL INFORMATION PROVIDED 01/22/20: IT WAS REPORTED DUE TO THE DELAY OF THE CASE, THE PATIENT HAD TO UNDERGO MORE ANESTHESIA. IT WAS ALSO CONFIRMED THAT THE CUT SUTURE WAS NOT FULLY RETRIEVED FROM THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 9693158 ↗
Other INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices
- Arthrex® AR-13995NArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE430 reports
- Arthrex® AR-12990NArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE197 reports
- Arthrex® AR-13991NArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE128 reports
- Arthrex® AR-8981-08SArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall3 reports
- Arthrex® AR-8981-06SArthrex, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall1 report
- Centurion EB16Centurion Medical Products Corporation · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall0 reports
- Medline Industries, Inc. DYNJ901391DMedline Industries, Inc. · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall0 reports
- HarmonyAIR LAS0042Steris Corporation · INSTRUMENT, MANUAL, SURGICAL, GENERAL USE1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page