Arthrex® AR-8973R-30-130Recalls & reported problems
Arthrex, Inc. · 3.0X130MM FIBULA NAIL, RIGHT · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00888867295216), received between Aug 5, 2020 and Sep 6, 2024 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Defective Device1
- Difficult to Remove1
- Failure to Advance1
- Material Twisted/Bent1
- Physical Resistance/Sticking1
Patient problems
- Insufficient Information3
- No Clinical Signs, Symptoms or Conditions2
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 6, 2024Failure to Advance; Failure to Advance
ON (B)(6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-8973R-30-130 ARTHREX RIGHT FIBULOCK NAIL TALONS WOULD NOT EXPAND OR ENGAGE WITH THE ACTUATION DRIVER. A NEW AR-8973R-30-130 ARTHREX RIGHT FIBULOCK NAIL WAS USED, AND THE ACTUATION DRIVER ENGAGED IMMEDIATELY. THE PATIENT WAS NOT AFFECTED. THIS WAS DISCOVERED DURING AN ANKLE FRACTURE PROCEDURE ON (B)(6) 2024. THE CASE WAS DELAYED THREE MINUTES WITHOUT ADDITIONAL ANESTHESIA ADMINISTERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20161842 ↗MalfunctionMay 2, 2024Defective Device; Defective Device
ON (B)(6) 2024, IT WAS REPORTED BY A FACILITY REPRESENTATIVE VIA EMAIL THAT AN AR-8973R-30-130 FIBULOCK NAIL PROXIMAL TALONS ON THE FIBULAR NAIL DID NOT DEPLOY WHEN ENGAGED WITH THE ACTUATOR. WHEN THE SURGEON PLACED THE ACTUATOR INTO THE NAIL, IT NEVER CLICKED OR CONNECTED WITH THE IMPLANT BUT INSTEAD JUST SPUN IN PLACE. THE SURGEON REMOVED THE NAIL FROM THE FIBULA TO SEE IF IT WAS PLACED PROPERLY ON THE JIG, AND IF ANYTHING WAS BLOCKING THE ACTUATOR. THE IMPLANT WAS TIGHTLY SECURED, AND THERE WAS STILL NO ENGAGEMENT. THE FIBULOCK NAIL WAS REMOVED AND ANOTHER AR-8973R-30-130 FIBULOCK NAIL WAS USED TO COMPLETE THE CASE.
Patient problems: Insufficient Information
MAUDE report 19229458 ↗InjuryAug 17, 2022Material Twisted/Bent
ON 7/28/2022 IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-8973R-30-130 FIBULA NAIL HAS NO 60 DEGREE BEND. THE NAIL WAS STRAIGHT. THIS WAS DISCOVERED DURING A FIBULA FRACTURE PROCEDURE. THE NAIL WAS IMPLANTED BUT DID NOT FIT BECAUSE THERE WAS NO BEND. THE SURGEON REMOVED IT AND CLOSED THE CASE WITHOUT HAVING FIXATION ON THE FIBULA SIDE.
Patient problems: Insufficient Information
MAUDE report 15242838 ↗MalfunctionJul 20, 2021Break
IT WAS REPORTED PER THE SALES REP DURING AN ANKLE FRACTURE PROCEDURE THE DOCTOR DID NOT REAM DOWN TO THE RECOMMENDED DEPTH AND DURING INSERTION THE TIP OF THE NAIL BROKE. ALL WAS RETRIEVED. THE SURGEON REAMED FURTHER AND IMPLANTED ANOTHER LONGER NAIL AND THERE WAS NO FURTHER ISSUES. THE PATIENT IS FINE TO DATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12196423 ↗InjuryJan 18, 2021Break; Physical Resistance/Sticking
IT WAS REPORTED, THAT DURING AN ANKLE FRACTURE REPAIR PROCEDURE, DURING INSERTION OF THE AR-8972R-30-130 THE PROXIMAL TALONS OPENED PRIOR TO DEPLOYMENT WITH THE DRIVER. THE SURGEON TRIED TO TAP WITH THE IMPACTOR HOWEVER, STILL A LOT OF RESISTANCE. THE SURGEON REMOVED THE AR-8972R-30-130, USED THE WHITE HANDLED DRIVER TO DISENGAGED THE TALONS, UNTIL THEY WERE FLUSH WITH THE SHAFT. THE SURGEON INSERTED THE SAME AR-8972-30-130 AND THE SAME PROBLEM OCCURRED. THE AR-8972R-30-130 WAS LOGGED AND THE TALONS BROKE OFF. THE SURGEON REMOVED THE AR-8972-30-130 UNABLE TO RETRIEVE THE BROKEN TALON. THE SURGEON USED THE SAME INSERTION SITE AND COMPLETED THE CASE USING A DIFFERENT AR-8972R-30-130.
Patient problems: Insufficient Information
MAUDE report 11186228 ↗InjuryAug 5, 2020Break; Difficult to Remove
IT WAS REPORTED DURING AN OPEN REDUCTION INTERNAL FIXATION ANKLE FRACTURE FIBULA REPAIR THE NAIL WAS INSERTED PER TECHNIQUE, THE SURGEON DID NOT LIKE THE POSITION SO HE TRIED TO REMOVE IT. DURING THE REMOVAL IT BROKE AND THE END CAP REMAINED IN THE PATIENT. IT WAS OBSERVED THE TALONS OF THE IMPLANT WERE NOT OPEN. THE DOCTOR REPLACED IT WITH ANOTHER NAIL AND IT WAS IMPLANTED WITH NO ISSUES BUT A PORTION OF THE FIRST NAIL STILL REMAINS IN THE PATIENT. THE PROCEDURE WAS COMPLETED. THE DOCTOR AND THE PATIENT ARE PLANNING TO DO A REVISION TO REMOVE THE FIRST PARTIAL NAIL.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 10366844 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page