Arthrex® AR-8973R-30-180Recalls & reported problems
Arthrex, Inc. · 3.0X180MM FIBULA NAIL, RIGHT · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867295230), received between Aug 10, 2023 and Jun 5, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Material Twisted/Bent1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Foreign Body In Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 5, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
ON 05/13/2025, AN ARTHREX SUBSIDIARY EMPLOYEE REPORTED VIA EMAIL THAT AN AR-8973R-30-180 ARTHREX FIBULOCK NAIL HAD AN ISSUE AFTER BEING INSERTED. DURING THE PROCEDURE, THE DEVICE WAS INSERTED, BUT WHEN ATTEMPTING TO INSERT THE PRODUCT A LITTLE DEEPER BEFORE OPENING THE TALON, THE TALON COULD NOT BE FULLY CLOSED. AS A RESULT, FURTHER INSERTION WAS STOPPED, AND THE DEVICE WAS REMOVED. A REPLACEMENT AR-8973R-30-180 ARTHREX FIBULOCK NAIL WAS USED INSTEAD. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22151310 ↗MalfunctionApr 18, 2025Break; Break
ON 03/24/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-8973R-30-180 ARTHREX FIBULOCK NAIL BROKE AT THE TIP OF WHERE THE TALONS DEPLOY DURING THE BEGINNING INSERTION OF THE NAIL. THE TALONS WERE COMPROMISED AND COULD NOT BE DEPLOYED PROPERLY. THERE WAS AN INCREASE IN CASE TIME, AND ANESTHESIA WAS INCREASED BY 15-20 MINUTES. THE BROKEN IMPLANT PORTION WAS LEFT INSIDE THE PATIENT. THE BROKEN PIECE WAS ADVANCED UP THE CANAL TO MAKE SURE THAT THE FINAL IMPLANT COULD BE USED. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Foreign Body In Patient
MAUDE report 21864679 ↗MalfunctionAug 10, 2023Material Twisted/Bent; Material Twisted/Bent
ON (B)(6) 2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-8973R-30-180 FIBULOCK® NAIL BENT OVER SIDEWAYS. THIS OCCURRED DURING AN OPEN TIBIAL FRACTURE ON (B)(6) 2023, AFTER REAMING THE NAIL WAS INSERTED AND AFTER INSERTING THE PLATE AND THE TIP OF THE NAIL WAS BENT OVER SIDEWAYS THE DEVICE WAS PULLED OUT AND THE CASE WAS COMPLETED USING A AR-8973R-30-130. THERE WAS NO PATIENT EFFECT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17497518 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page