Arthrex® AR-8973L-38-180Recalls & reported problems
Arthrex, Inc. · 3.8X180MM FIBULA NAIL, LEFT · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867295261), received between Feb 15, 2019 and Oct 24, 2024 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Failure to Advance1
- Patient Device Interaction Problem1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
- Pain1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 24, 2024Failure to Advance; Failure to Advance
ADDITIONAL INFORMATION RECEIVED ON 10/23/2024: THIS WAS DISCOVERED WHILE THE NAIL WAS ALREADY IN THE PATIENT. NOTHING BROKE OFF IN THE PATIENT. THE CASE WAS COMPLETED USING AN AR-8973L-30-180 ARTHREX FIBULOCK NAIL, A SMALLER NAIL. THE CASE WAS ONLY DELAYED LONG ENOUGH TO GET THE NAIL. ADDITIONAL ANESTHESIA WAS NOT NEEDED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 20522113 ↗InjurySep 6, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
ON (B)(6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT A PATIENT UNDERWENT REVISION SURGERY DUE TO PAIN. ON (B)(6) 2023, THE PATIENT UNDERWENT SURGERY FOR AN ANKLE FRACTURE AND SYNDESMOSIS REPAIR. DURING THE PROCEDURE, TWO AR-8730-36PT LOW PROFILE SCREW, TI, 3.0 MM X 36 MM, CANCELLOUS, PARTIALLY THREADED, AN AR-8827-18 LOW PROFILE SCREW, 2.7 X 18 MM, CORTEX, AN AR-8827-22 LOW PROFILE SCREW, 2.7 X 22 MM, CORTEX, AN AR-8973DS FIBULOCK IMPLANT SYSTEM, AN AR-8973L-38-180 ARTHREX FIBULOCK NAIL, 3.8 X 180 MM LEFT, AND AN AR-8925SS SYNDESMOSIS TIGHTROPE XP, STAINLESS STEEL WERE IMPLANTED. POST-OPERATIVE, THE PATIENT STARTED TO EXPERIENCE PAIN. ON (B)(6) 2024, REVISION SURGERY WAS PERFORMED TO REMOVE ALL THE IMPLANTS EXCEPT FOR THE AR-8973L-38-180 ARTHREX FIBULOCK NAIL. DURING THE REMOVAL OF THE IMPLANTS, THE DISTAL END OF THE EXTRACTION TOOL OF AN AR-8973RK FIBULOCK REMOVAL KIT BROKE OFF AFTER ROUGHLY 8 MALLET STRIKES. ANOTHER AR-8973RK FIBULOCK REMOVAL KIT WAS OPENED, AND THE SAME THING OCCURRED. THE EXTRACTION TOOLS BROKE, BUT NO PIECES WERE IN THE PATIENT. THE SURGEON ELECTED TO CLOSE AND FINISH THE CASE WITH THE NAIL STILL IMPLANTED AND FELT CONFIDENT THAT THE REMOVE
Patient problems: Pain
MAUDE report 20156946 ↗InjuryFeb 15, 2019Break
IT WAS REPORTED THAT DURING A ANKLE ARTHROSCOPY PROCEDURE, AFTER THE SURGEON ADVANCED THE FIBULA NAIL, AR-8973L-38-180, INTO THE DISTAL PORTION OF THE FIBULA, THE SURGEON HAD TO PULL THE NAIL BACK TO RE-ALIGN THE DISTAL PORTION WITH THE PROXIMAL PORTION OF THE FIBULA. WHEN HE RETRACTED THE NAIL BACK THE HEAD OF THE NAIL BROKE. THE CASE WAS COMPLETED BY PLATING AFTER THE INCIDENT. *ADDITIONAL INFORMATION REQUESTED. ADDITIONAL INFORMATION PROVIDED 1/25/2019: THIS WAS A FIBULA FRACTURE REPAIR PROCEDURE. THE TIP OF THE NAIL BROKE BUT DID NOT DETACH FROM THE REST OF THE DEVICE; THEREFORE, NO FRAGMENTS REMAINED IN THE PATIENT. AN ADDITIONAL INCISION WAS CREATED FOR THE INSERTION OF THE PLATE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8344553 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.