Arthrex® AR-8973R-38-180Recalls & reported problems

Arthrex, Inc. · 3.8X180MM FIBULA NAIL, RIGHT · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867295278), received between Oct 25, 2023 and Oct 25, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Defective Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Defective Device1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 25, 2023Defective Device; Defective Device

    ON 9/28/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-8973R-38-180 ARTHREX FIBULOCK NAIL TALONS DID NOT DEPLOY AND THE ACTIVATION DRIVER WAS NOT SEATING ALL THE WAY. THE SURGEON REMOVED THE ARTHREX FIBULOCK NAIL AND ATTEMPTED TO ACTIVATE THE TALONS OFF THE JIG BUT WAS UNSUCCESSFUL. ANOTHER AR-8973R-38-180 ARTHREX FIBULOCK NAIL WAS USED TO COMPLETE THE CASE. THIS WAS DISCOVERED DURING A FIBULOCK NAIL PROCEDURE ON 9/26/2023. AFTER THE PROCEDURE, THE SALES REPRESENTATIVE PUT THE FIBULOCK NAIL THROUGH THE WASH AND SPENT ABOUT TEN TO FIFTEEN MINUTES TRYING TO DEPLOY THE TALONS. ONCE THE TALONS HAD DEPLOYED, THE SALES REPRESENTATIVE NOTICED THAT THERE WAS A SMALL GRAY BAR OR LEDGE INTERFERING WITH THE DRIVER.

    Patient problems: Insufficient Information

    MAUDE report 18005378 ↗

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page