Arthrex® AR-9090-02SRecalls & reported problems

Arthrex, Inc. · DualComp Hindfoot Nail, 10.5 x 210 mm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867305502), received between Jun 21, 2024 and Sep 16, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Difficult to Advance1
  • Loose or Intermittent Connection1
  • Positioning Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 16, 2025Positioning Failure; Positioning Failure

    ON 08/26/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-9090-02S DUALCOMP HINDFOOT NAIL DID NOT ALIGN PROPERLY WITH THE GUIDED ARM AND JIG. THIS WAS DISCOVERED DURING THE CASE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23064502 ↗
  • MalfunctionApr 23, 2025Loose or Intermittent Connection; Loose or Intermittent Connection

    ON 04/10/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA PHONE THAT AN AR-9090-02S DUALCOMP HINDFOOT NAIL NITINOL CORD BECAME LOOSE. THIS OCCURRED DURING HINDFOOT FUSION ON (B)(6) 2025, WHEN THEY GOT THE IMPLANT IN AND THE NITINOL CORD BECAME LOOSE OR DID NOT WORK. EVERYTHING WAS REMOVED. THE CASE CONTINUED USING ANOTHER NAIL. THERE WAS A DELAY OF ABOUT 30 MINUTES AND IT IS UNKNOWN IF ADDITIONAL ANESTHESIA WAS ADMINISTERED. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21895599 ↗
  • MalfunctionJun 21, 2024Break; Difficult to Advance; Break; Difficult to Advance

    ON 05/24/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT THE PEEK PIECE HOLDING THE SLIDING INTERNAL COMPONENT OF AN AR-9090-02S DUAL COMP HINDFOOT NAIL BROKE AND ENGAGED DURING INSERTION. THIS CAUSED SLACK WITHIN THE TENSION WIRE BEFORE THE NAIL WAS COMPLETELY INSERTED. THE SURGEON CLEARED THE ANKLE OF THE CARTILAGE AND CORRECTED DEFORMITIES. PLACED TWO 2.4 MM GUIDEWIRES TO HOLD THE ANKLE IN POSITION. PLACED 3MM STARTING GUIDE PIN AND REAMED WITH THE OPENING REAMER UNTIL THE END OF THE FLUTES HAD ENTERED THE CALCANEUS. THE 3MM WIRE WAS REMOVED AND THE BALL TIP WIRE WAS PLACED INTO THE CANAL. THE SURGEON STARTED WITH THE 9MM FLEXIBLE REAMER, INCREASING BY 1MM UNTIL REACHING THE FINAL 12MM REAMER. THEN, SELECTED AN AR-9090-02S DUALCOMP HINDFOOT NAIL, 10.5MM X 210MM, AND LOADED IT AS THE TECHNIQUE GUIDE EXPLAINS. THE SURGEON PACKED ANKLE JOINTS WITH AUTOGRAFT AND ALLOGRAFT BONE. THEN, THE SURGEON BEGAN TO INTRODUCE THE NAIL, WITH IT BECOMING DIFFICULT TO ADVANCE WITHIN THE FIRST 25 PERCENT OF INSERTION. THE STRIKE PLATE WAS ATTACHED, AND THE SURGEON BEGAN TO MALLET THE NAIL. ONCE THE NAIL HAD ADVANCED TO ABOUT HALFWAY, IT WAS NOTICED THAT THE TENSIONING CABLE HAD SLACK

    Patient problems: Insufficient Information

    MAUDE report 19583683 ↗