Arthrex® AR-9090-04SRecalls & reported problems

Arthrex, Inc. · DualComp Hindfoot Nail, 10.5 x 300 mm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867305526), received between Mar 15, 2024 and Mar 15, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Failure to Advance” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Advance1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 15, 2024Failure to Advance; Failure to Advance

    ON 02/29/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA E-MAIL THAT AN AR-9090-04S DUALCOMP HINDFOOT NAIL COULD NOT BE IMPLANTED. THIS OCCURRED DURING A DUAL COMPRESSION TTC NAIL CASE WHEN TENSIONING THE CABLE FOR THE COMPRESSION MECHANISM; THE HANDLE MAXED OUT IN THE TENSIONER. WHEN THEY WENT TO DRILL THE PROXIMAL CALCANEAL SCREW (THIRD SCREW INTO THE NAIL), THE TRAJECTORY THROUGH THE JIG WAS BLOCKED BY THE INTERNAL SLIDING MECHANISM OF THE NAIL. IT WAS AT THIS POINT THAT THEY REALIZED THE INTERNAL SLIDER HAD NOT REACHED ITS MAXIMUM DISTANCE. THEY TRIED TO TROUBLESHOOT ON THE FLY, SEEING IF THEY COULD FORCE THE SLIDER DOWN THROUGH A SCREW HOLE OR SEE IF ANYTHING WAS BLOCKING THE MECHANISM, BUT WERE UNABLE TO MOVE THE SLIDER.

    Patient problems: Insufficient Information

    MAUDE report 18916338 ↗

Other Rod, Fixation, Intramedullary And Accessories devices

All Rod, Fixation, Intramedullary And Accessories devices →