Skip to content

Arthrex® AR-9090-11SRecalls & reported problems

Arthrex, Inc. · DualComp Hindfoot Nail, 12.5 x 240 mm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867305595), received between Feb 26, 2025 and Feb 26, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Positioning Failure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Positioning Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 26, 2025Positioning Failure; Positioning Failure

    ON 02/13/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-9090-11S DUALCOMP HINDFOOT NAIL HAD AN ISSUE WHEN TENSIONING THE NAIL ON STEP 3 ON THE JIG. THE COMPRESSION DEVICE BOLTS WERE "BOTTOMED OUT" DURING TENSIONING AND "ZEROED OUT" BEFORE INSERTION. HOWEVER, THE NITINOL CORE IN THE NAIL DID NOT FULLY OPEN UP THE TALAR SCREW OR THE SUPERIOR CALC SCREW. THEY HAD TO REMOVE THE NAIL AND USE A DIFFERENT ONE AS THE SURGEON WAS CONCERNED ABOUT NOT HAVING ENOUGH POINTS OF FIXATION DUE TO THE ISSUE. THEY BELIEVE THE NITINOL CORE BEGAN TO MOVE AND THEN GOT STUCK AND WOULD NOT OPEN UP THE TALAR SCREW OR SUPERIOR CALC SCREW ANYMORE. THEREFORE, THEY HAD TO REMOVE AND REPLACE THE NAIL WITH ANOTHER NAIL WITHOUT ISSUES. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2025, WITH NO PATIENT HARM REPORTED. ADDITIONAL INFORMATION WAS RECEIVED ON 02/18/2025: NO REPORTED PIECES OF THE AR-9090-11S DUALCOMP HINDFOOT NAIL BROKE OFF INSIDE THE PATIENT. THE PROCEDURE WAS DELAYED BY 45 MINUTES, AND BECAUSE IT TOOK AN ADDITIONAL 45 MINUTES, ADDITIONAL ANESTHESIA WAS ADMINISTERED, BUT THERE IS NO REPORT OF THE EXACT QUANTITY ADMINISTERED. THE CASE WAS COMPLETED USING A DIFFERENT ARTHREX

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21470004 ↗

Other Rod, Fixation, Intramedullary And Accessories devices

All Rod, Fixation, Intramedullary And Accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page