Arthrex® AR-8970JDRecalls & reported problems

Arthrex, Inc. · Mini Joint Distractor/Compressor · Class 1

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00888867307377), received between Apr 18, 2024 and Mar 24, 2026 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Physical Resistance/Sticking” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 3 reportsJun 2024: 3 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Physical Resistance/Sticking4
  • Break1
  • Defective Device1
  • Difficult to Open or Close1
  • Mechanical Jam1
  • Positioning Failure1
  • Retraction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • Insufficient Information2

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 24, 2026Positioning Failure

    ON 4-MAR-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-8970JD JOINT DISTRACTOR WILL DEPRESS BUT NOT DISTRACT. NOTICED DURING A CASE, DEVICE SWAPPED, AND CASE COMPLETED WITH NO PATIENT EFFECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24673172 ↗
  • MalfunctionNov 5, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    ON 10/27/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-8970JD MINI JOINT DISTRACTOR/COMPRESSOR HAD A WIRE COLD WELD INTO THE PIN DISTRACTOR. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23475161 ↗
  • MalfunctionJan 29, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    ON 01/06/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-8970JD MINI JOINT DISTRACTOR/COMPRESSORS SWITCH TO FLIP THE DISTRACTOR FROM COMPRESSION-NEUTRAL-DISTRACTION POSITIONS, IS DIFFICULT TO ENGAGE. THIS OCCURRED DURING A CASE AND CAUSE NO HARM ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21264012 ↗
  • MalfunctionJun 19, 2024Break; Break

    IT WAS REPORTED THAT THE AR-8970JD KNOB BROKE FOLLOWING PLACEMENT OF TWO 2.4MM GUIDE WIRES FOR TALONAVICULAR JOINT DISTRACTION WITH DEBRIDEMENT. THE KNOB WAS DROPPED OUTSIDE OF THE PATIENT, GUIDE WIRES WERE REMOVED AND A SMALL JOINT DISTRACTOR WAS USED IN ITS PLACE. THE TENSIONING KNOB WAS RETRIEVED FROM THE OPERATING ROOM FLOOR FOLLOWING CASE COMPLETION.

    Patient problems: Insufficient Information

    MAUDE report 19566832 ↗
  • MalfunctionJun 19, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT ONE OF THE ARMS ON THE DISTRACTOR IS LOCKING UP AND NOT WORKING CORRECTLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19566529 ↗
  • MalfunctionJun 13, 2024Mechanical Jam; Mechanical Jam

    ON 10/13/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT A AR-8970JD MINI JOINT DISTRACTOR/COMPRESSOR BECAME JAMMED. THIS OCCURRED DURING AN UNKNOWN CASE ON (B)(6) 2022, THE CASE CONTINUED, WITH NO PATIENT EFFECT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19527008 ↗
  • MalfunctionMay 20, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    ON 04/13/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-8970JD MINI JOINT DISTRACTOR/COMPRESSOR WOULDN'T UNLOCK PROPERLY AFTER BEING DISTRACTED. THIS WAS DISCOVERED DURING USE WITH NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19351154 ↗
  • MalfunctionMay 14, 2024Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED ON 04/10/2022 BY A FACILITY REPRESENTATIVE VIA SEMS THAT AN AR-8970JD JOINT DISTRACTOR IS STUCK. THIS WAS DISCOVERED DURING A CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19308744 ↗
  • MalfunctionMay 7, 2024Retraction Problem; Retraction Problem

    IT WAS REPORTED THAT DURING A SURGERY THERE WERE TWO PROBLEMS. THE SCREW DRIVER FOR 4,5 MM SCREWS WITH THE GOLDEN TIP DIDN´T FIT IN THE CORTICAL SCREW AR-8545-XX PROPERLY. AND THE MINI JOINTS DISTRACTOR AR-8970JD JAMMED AND ONLY COULD EXTRACT, BUT NOT COMPRESS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19261482 ↗
  • MalfunctionApr 18, 2024Defective Device; Defective Device

    ON 03/22/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS-06522233 THAT AN AR-8970JD MINI JOINT DISTRACTOR/COMPRESSOR FAILED DURING A CASE. NO ADDITIONAL INFORMATION WAS PROVIDED, AND ADDITIONAL INFORMATION HAS BEEN ASKED.

    Patient problems: Insufficient Information

    MAUDE report 19130935 ↗

Other Apparatus, traction, non-powered devices