NA 72201725Recalls & reported problems
Smith & Nephew, Inc. · SUPINE TABLE ATTACHMENT WITH PAD · Class 1
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 03596010612816), received between Mar 27, 2018 and Oct 17, 2023 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Material Separation2
- Positioning Failure2
- Detachment Of Device Component1
- Positioning Problem1
- Unstable1
Patient problems
- No Known Impact Or Consequence To Patient5
- No Clinical Signs, Symptoms or Conditions2
- Fall1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 17, 2023Break; Break
IT WAS REPORTED THAT DURING A PROCEDURE, THE SUPINE TABLE BROKE WHILE PULLING TRACTION. THE PROCEDURE COULD NOT BE COMPLETED AND WAS RESCHEDULED. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17953868 ↗MalfunctionApr 14, 2021Unstable; Unstable
IT WAS REPORTED THAT THE SUPINE TABLE ATTACHMENT DID NOT LOCK AWAY PROPERLY DURING SET UP FOR A HIP SCOPE. THE SCREW WAS STRIPED SO THE ATTACHMENT COULD NOT BE SECURELY ATTACHED TO THE BED. THERE WAS A SMITH AND NEPHEW BACK UP DEVICE. NO PATIENT INJURY, DELAYS OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11663425 ↗MalfunctionSep 17, 2020Positioning Problem; Positioning Problem
IT WAS REPORTED THAT DURING ARTHROSCOPY, THE LEG HOLDER DETACHED FROM THE TABLE ATTACHMENT WHEN TRACTION WAS APPLIED TO THE OPERATIVE LEG, DESPITE ALL REQUIRED CHECKS BEING CARRIED OUT PRIOR TO SURGERY. THERE WAS A DELAY OF UNDER 30 MIN AND THE PROCEDURE WAS COMPLETED USING THE SAME DEVICE AND STANDARD SUPINE TECHNIQUE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10547153 ↗MalfunctionMar 10, 2020Positioning Failure; Material Separation; Positioning Failure; Material Separation
IT WAS REPORTED THAT DURING A HIP ARTHROSCOPY, THE STRAP WHICH SECURES THE FOOT INTO THE ACTIVE HEEL TRACTION BOOT BROKE SINCE IT WAS ROUTINELY OVER TIGHTEN. SINCE THERE WAS NOT ANY BACK-UP DEVICE AVAILABLE, IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED; NEVERTHELESS, THE PROCEDURE WAS NOT DELAYED AND NO PATIENT INJURIES WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9814132 ↗MalfunctionMar 9, 2020Material Separation; Material Separation
IT WAS REPORTED THAT DURING A HIP ARTHROSCOPY, THE STRAP WHICH SECURES THE FOOT INTO THE ACTIVE HEEL TRACTION BOOT BROKE SINCE IT WAS ROUTINELY OVER TIGHTEN. SINCE THERE WAS NOT ANY BACK-UP DEVICE AVAILABLE, IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED; NEVERTHELESS, THE PROCEDURE WAS NOT DELAYED AND NO PATIENT INJURIES WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9808750 ↗MalfunctionAug 26, 2019Positioning Failure; Positioning Failure
IT WAS REPORTED THAT THE OPERATIVE LEG HAS COME OUT OF THE SUPINE TABLE ATTACHMENT MID PROCEDURE WHEN THE SURGEON WAS MOVING THE LEG BACK INTO EXTENSION DURING THE OSTEOPLASTY. A KNOT WAS NOT TIED UP WHEN THE PROCEDURE WAS TAKING PLACE. A LOSS OF TRACTION WAS REPORTED DURING USE. PROCEDURE WAS COMPLETED WITH THE SAME DEVICE WITH NO DELAY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8930138 ↗MalfunctionMar 11, 2019Break
IT WAS REPORTED THAT THE PLATE BROKE OFF. THERE WAS PATIENT INVOLVED, THE SAME DEVICE WAS ABLE TO BE USED AND THE PROCEDURE WAS DELAYED FOR MORE THAN 60 MINUTES. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8410461 ↗MalfunctionMar 27, 2018Detachment Of Device Component
IT WAS REPORTED DURING SET-UP FOR A HIP POSITIONING SYSTEM, THE UNIVERSAL HIP DISTRACTOR WERE ATTACHED TO THE TABLE EXTENSION, THE OPERATOR FULLY INSERTED INTO THE SOCKET AND TIGHTENED THE HOLDER KNOB. AFTER SETTING-UP THEM, TO APPLY TENSION TO THE WELL LEG, WHEN THE LEG HOLDER WAS PULLED, THE HIP DISTRACTOR CAME OFF FROM THE TABLE EXTENSION AND THE PATIENT LEGS FALL OFF. THIS HAPPENED DURING PREPARATION OF THE PROCEDURE, NO SURGICAL INSTRUMENTS WERE INSIDE THE PATIENT. THERE WAS NO REPORT OF PROCEDURAL DELAY. NO PATIENT INJURIES OR COMPLICATIONS WERE REPORTED. A COMPETITOR DEVICE WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: Fall
MAUDE report 7375414 ↗
Other APPARATUS, TRACTION, NON-POWERED devices
- Arthrex® AR-1600MArthrex, Inc. · Apparatus, traction, non-powered21 reports
- Active Heel Traction Boot 72203270Smith & Nephew, Inc. · APPARATUS, TRACTION, NON-POWERED20 reports
- NA 357.371Synthes GmbH · APPARATUS, TRACTION, NON-POWERED16 reports
- NA SD393.640Synthes GmbH · APPARATUS, TRACTION, NON-POWERED15 reports
- NA 394.44Synthes GmbH · APPARATUS, TRACTION, NON-POWERED12 reports
- Active Heel Traction Boot 72202682Smith & Nephew, Inc. · APPARATUS, TRACTION, NON-POWERED10 reports
- Arthrex® AR-8970JDArthrex, Inc. · Apparatus, traction, non-powered10 reports
- Pivot Guardian 3105000100Stryker Corporation · Apparatus, traction, non-powered6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page