Skip to content

Arthrex® AR-8656-02Recalls & reported problems

Arthrex, Inc. · CHONDRAL PICK, 60 DEG · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867307452), received between Apr 12, 2021 and May 24, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent3
  • Flaked1
  • Material Fragmentation1

Patient problems

  • Insufficient Information3
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 24, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE SURGEON WAS ATTEMPTING TO USE THE CHONDRAL PICKS TO PERFORM A MICRO FRACTURE ON PATIENT'S TALUS. PICKS WOULD NOT PENETRATE BONE AND JUST BENT OUT OF THE WAY. K-WIRES WERE USED TO COMPLETE THE CASE.

    Patient problems: Insufficient Information

    MAUDE report 19391274 ↗
  • MalfunctionJan 16, 2024Material Twisted/Bent; Material Twisted/Bent

    ON 12/18/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS (B)(4) THAT AN AR-8656-02 CHONDRAL PICK BENT. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2023.

    Patient problems: Insufficient Information

    MAUDE report 18519235 ↗
  • MalfunctionNov 6, 2023Material Fragmentation; Material Fragmentation

    ON 10/10/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-8656-02 CHONDRAL PICK AND AN AR-8656-03 CHONDRAL PICK WERE PRODUCING SMALL METAL SHAVINGS. THIS OCCURRED DURING A CASE WHERE THE DEBRIS WAS RETRIEVED. THE DEVICES WERE SWITCHED OUT AND THE CASE WAS COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18075496 ↗
  • MalfunctionApr 12, 2021Flaked; Material Twisted/Bent

    IT WAS REPORTED THAT DURING USE IN AN ANKLE ARTHROSCOPY, THE AR-8656-02 PICK BENT AND FLAKES WENT INTO THE PATIENT. THE FLAKES WERE REMOVED WITH SHAVER SUCTION.

    Patient problems: Insufficient Information

    MAUDE report 11648345 ↗

Other INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices

All INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices →

Sources and updates