Arthrex® AR-5400-02Recalls & reported problems
Arthrex, Inc. · GLENOID TARGETER HANDLE · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00888867315341), received between May 22, 2024 and Mar 21, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Material Twisted/Bent” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Twisted/Bent2
- Defective Device1
- Fitting Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Insufficient Information1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 21, 2025Defective Device; Defective Device
CLARIFICATION: IT WAS REPORTED THAT THE TARGETER ARMS, AR-5400-10, AR-5400-12, AR-5400-15, AR-5400-18, AND AR-5400-25 WERE SLIDING ON THE HANDLE, DESPITE BOTH NUTS BEING SECURELY TIGHTENED.
Patient problems: Insufficient Information
MAUDE report 21662875 ↗MalfunctionMar 7, 2025Material Twisted/Bent; Material Twisted/Bent
ON 02/15/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-5400-02 TARGETING HANDLE PRONGS ARE BENT AND WILL NOT SEAT. THIS OCCURRED DURING USE IN A CASE WITH NO PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21549903 ↗MalfunctionJun 4, 2024Material Twisted/Bent; Material Twisted/Bent
ON 2/3/2023, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS THAT AN AR-5400-02 GLENOID TARGETER HANDLE HAD AN ISSUE. THE TOP PART OF THE ITEM WAS LOOSE AND SEEMED TO BEND WHEN TRYING TO ENGAGE. THEY HAD TO OPEN UP ANOTHER PAN AND USE THE ONE FROM IT. THIS WAS DISCOVERED DURING USE WITH NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19459367 ↗MalfunctionMay 22, 2024Fitting Problem; Fitting Problem
ON 08/19/2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-5400-02 GLENOID TARGETER HANDLE AND AN AR-5400-01 GLENOID TARGETER SHAFT HAD AN ISSUE. DURING AN UNSPECIFIED PROCEDURE, THE AR-5400-01 GLENOID TARGETER SHAFT SLOT A WOULD NOT ALLOW THE ARM TO SLIDE IN, AND THE AR-5400-02 GLENOID TARGETER HANDLE WOULD NOT ALLOW THE ARM TO SLIDE INTO THE A SLOT. THIS WAS DISCOVERED DURING USE WITH NO IMPACT TO THE PATIENT. ADDITIONAL INFORMATION WAS REQUESTED. ON 08/31/2022, THE SALES REPRESENTATIVE STATED THE FOLLOWING INFORMATION VIA PHONE: THIS EVENT OCCURRED DURING A REVERSE TOTAL SHOULDER PROCEDURE ON (B)(6) 2022. THE SURGEON USED A HEMOSTAT TO GUIDE THE DEVICE INTO THE A SLOT, AND BOTH DEVICES THEN WERE USED TO COMPLETE THE CASE SUCCESSFULLY WITH NO IMPACT TO THE PATIENT. THESE INSTRUMENTS COULD NOT BE USED ANYMORE AFTER THE COMPLETION OF THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19370506 ↗
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page