Arthrex® AR-5400-15Recalls & reported problems
Arthrex, Inc. · SIZE 15 LEG, GLENOID TARGETER · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867315426), received between Feb 5, 2025 and Oct 29, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Defective Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Defective Device1
- Material Twisted/Bent1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 29, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
ON 10/08/2025, A FACILITY REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-5400-15 VIP GLENOID TARGETER LEG WOULD NOT SLIDE INTO THE TARGETER DURING A CASE WITH NO PATIENT AFFECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23412605 ↗MalfunctionMar 26, 2025Defective Device; Defective Device
ON 03/03/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-5400-02 VIP¿ GLENOID TARGETER HANDLE SLID EVEN THOUGH BOTH NUTS WHERE SECURED TIGHTLY. AR-5400-04 VIP¿ GLENOID TARGETER LOCKING NUT, AND A SECOND NUT FROM THE AR-4500-S SET. THIS OCCURRED DURING A REVERSE SHOULDER ARTHROPLASTY ON (B)(6) 2025. ADDITIONAL INFORMATION WAS PROVIDED ON 03/04/2025 WHERE THE SALES REPRESENTATIVE STATED THE ARMS THAT DIDN¿T HOLD IN PLACE WERE 10,12,15,18,25. CLARIFICATION: IT WAS REPORTED THAT THE TARGETER ARMS, AR-5400-10, AR-5400-12, AR-5400-15, AR-5400-18, AND AR-5400-25 WERE SLIDING ON THE HANDLE, DESPITE BOTH NUTS BEING SECURELY TIGHTENED.
Patient problems: Insufficient Information
MAUDE report 21696580 ↗MalfunctionFeb 5, 2025Material Twisted/Bent; Material Twisted/Bent
ON 01/10/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4). THAT AN AR-5400-15 VIP¿ GLENOID TARGETER, SIZE 15 LEG WAS BENT DURING A CASE. THIS WAS DISCOVERED DURING A PROCEDURE, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION WAS RECEIVED ON 1/13/2025: THIS WAS DISCOVERED DURING A REVERSE SHOULDER PROCEDURE ON 12/30/2024. AN EXTRA GUIDE WAS USED TO COMPLETE THE CASE. THERE WAS NO CASE DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21315276 ↗
Other System, image processing, radiological devices
- Interventional Workspot Interventional Workspot - 1Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological13 recalls24 reports
- Synapse Radiology PACS Synapse Radiology PACS Software LicenseFujifilm Healthcare Americas Corporation · System, image processing, radiological9 recalls38 reports
- CENTRICITY UNIVERSAL VIEWER ZFP CLIENT 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 7.0Ge Healthcare · System, image processing, radiological6 recalls0 reports
- IntelliSpace Cardiovascular 8Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological2 recalls132 reports
- CA1000 2.0 SP 14Ge Healthcare · System, image processing, radiological4 recalls0 reports
- VMA 2Ortho Kinematics, Inc. · System, Image Processing, Radiological4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page