Skip to content

Arthrex® AR-8717G-02Recalls & reported problems

Arthrex, Inc. · DynaNite® Staple Drill Guide, 13 / 15 mm · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867326101), received between Apr 25, 2024 and May 24, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Physical Resistance/Sticking” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Physical Resistance/Sticking3

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 24, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    ON (B)(6) 2022, IT WAS REPORTED BY A DISTRIBUTOR VIA PHONE THAT ON (B)(6) 2022 DURING A 4 CORNER FUSION A AR-8717DS-1110 DYNANITE STAPLE IMPLANT SYSTEMS DRILL BIT GOT BOUND UP IN THE DRILL GUIDE. THEN FOLLOWING THIS AN AR-8717DS-0910 DYNANITE STAPLE IMPLANT SYSTEMS DRILL BIT ALSO GOT BOUND UP IN THE DRILL GUIDE. THE SURGEON PROCEEDED TO GET 2 NEW KITS AND THOSE WORKED ACCORDINGLY. A AR-8717G-01 DRILL GUIDE, AND A AR-8717G-02 DRILL GUIDE ALSO GOT BOUND UP. THE CASE WAS COMPLETED HOWEVER THE SURGEON WAS NOT PLEASED WITH THE PLACEMENT.

    Patient problems: Insufficient Information

    MAUDE report 19397566 ↗
  • MalfunctionMay 15, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    ON 8/19/2021 IT WAS REPORTED BY SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-8717D-02 GOT STUCK IN THE DRILL GUIDE AR-8717G-02 SIDE "15" DURING A CASE. THIS OCCURRED AGAIN DURING ANOTHER PROCEDURE, WHERE AR-8717D-02 GOT STUCK IN THE DRILL GUIDE AR-8717G-02 SIDE "13". NO PATIENT AFFECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19319259 ↗
  • MalfunctionApr 25, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT WHEN DRILLING FROM THE GUIDE SLEEVE PART, IT DID NOT PROCEED SMOOTHLY. THE CONTACT AREA AT THE BASE OF THE DRILL HAD A SCORCHED MARK. A NEW GUIDE AND DRILL WAS OPENED AND USED TO COMPLETE THE CASE WITH NO ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19185872 ↗

Other INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices

All INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page