Arthrex® AR-9091-02Recalls & reported problems
Arthrex, Inc. · Hindfoot Nail Cable Tensioner · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00888867335301), received between Mar 28, 2024 and Jul 25, 2025 — classified as 3 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Physical Resistance/Sticking” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Physical Resistance/Sticking4
- Defective Device1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Insufficient Information3
- Pain1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 25, 2025Physical Resistance/Sticking; Physical Resistance/Sticking
ON 06/30/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-9091-02 HINDFOOT NAIL CABLE TENSIONER CABLE SETTING SCREWS IN THE TENSIONER HAVE BEEN REVERSE THREADED INTO THE JIG AND FREQUENTLY GETS STUCK. THIS HAPPENED DURING A CASE AND HAD NO ADVERSE EFFECTS ON THE PATIENT. ADDITIONAL INFORMATION WAS PROVIDED VIA PHONE, WHERE THE SALES REPRESENTATIVE STATED THIS EVENT OCCURRED DURING AN ANKLE FUSION PROCEDURE ON (B)(6) 2025, WHERE THEY WERE UNABLE TO COMPRESS THE NAIL. THIS RESULTED IN A DELAY OF ABOUT 15-20 MINUTES THAT DID NOT REQUIRE ADDITIONAL ANESTHESIA
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22608001 ↗InjuryJan 6, 2025Defective Device; Physical Resistance/Sticking; Defective Device; Physical Resistance/Sticking
ON 12/9/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN ARTHREX LOANER TRAY WAS RECEIVED A COUPLE OF DAYS BEFORE THE SURGERY TOOK PLACE ON (B)(6) 2024. IT WAS VERIFIED THAT THE 9091-02 HINDFOOT NAIL CABLE TENSIONER WAS IN A ZEROED-OUT POSITION BEFORE BEING STERILIZED. ABOUT THIRTY MINUTES BEFORE THE SURGERY, IT WAS VERIFIED AGAIN THAT THE 9091-02 HINDFOOT NAIL CABLE TENSIONER WAS IN A ZEROED-OUT POSITION. DURING THE PROCEDURE, WHEN THE TECHNICIAN WENT TO TIGHTEN DOWN THE SET SCREW WITH THE HEX DRIVER, THE DRIVER IMMEDIATELY BEGAN TORQUING WITHOUT ACTUALLY TIGHTENING THE SCREW ITSELF. THE TECHNICIAN TRIED TO LOOSEN THE SET SCREW, BUT IT WAS PROVING VERY DIFFICULT TO LEFTY LOOSEY THE SCREW EVEN AFTER THE FIRST COUPLE OF TURNS. IT WAS ATTEMPTED AGAIN TO TIGHTEN THE SCREW DOWN AGAIN NOW, AND IT IMMEDIATELY CLICKED DUE TO THE TORQUE LIMITING FEATURE ON THE HEX DRIVER LIKE LAST TIME. THE CABLE TENSIONER WAS ALSO ABLE TO BE PULLED FREELY ON THAT SIDE DESPITE THE TORQUE LIMITER CLICKING LIKE IT WAS FULLY SEATED. THE OTHER SIDE WAS ATTEMPTED TO BE TIGHTENED, BUT THE SAME ISSUE WAS HAPPENING. AFTER ABOUT FIVE MINUTES, THE SURGEON NOTICED THE SET SCREWS WERE VERY ELEVATED AN
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 21079015 ↗InjuryOct 31, 2024Physical Resistance/Sticking; Physical Resistance/Sticking
ON 10/15/2024, A SALES REPRESENTATIVE REPORTED VIA SEMS-06686918 THAT AN AR-9091-02 HINDFOOT NAIL CABLE TENSIONER FAILED DURING A CASE. THE REP STATED THAT THE PATIENT WAS AFFECTED BY THE FAILURE. THEY COULD NOT USE THE NAIL'S TENSIONING COMPONENT AND HAD TO ADD ADDITIONAL HARDWARE TO COMPENSATE. THIS WAS DISCOVERED DURING A PROCEDURE. ADDITIONAL INFORMATION WAS RECEIVED ON 10/21/2024: THIS WAS DISCOVERED DURING A TTC NAIL PROCEDURE ON 10/14/2024. IT WAS STATED THAT THE FAILURE OF THE AR-9091-02 HINDFOOT NAIL CABLE TENSIONER WAS DUE TO THE SET SCREWS NOT MOVING UP AND DOWN WITHIN THE TENSIONER. THE SCREWS THEN ENDED UP SHEARING THE INNER WALL OF THE TENSIONER. THE SCREWS MAY HAVE ALSO BEEN STRIPPED. THE TENSIONER COULD NOT PULL THE NITINOL CORE OF THE NAIL TO GAIN COMPRESSION. THIS LEFT THE NAIL STATIC AND UNABLE TO USE THE SECOND CALCANEUS SCREW. THE CASE WAS COMPLETED SUCCESSFULLY BY USING ADDITIONAL FIXATION OUTSIDE OF THE NAIL. THE TECHNIQUE WAS CHANGED DUE TO NOT BEING ABLE TO USE THE COMPRESSION FEATURE OF THE NAIL, AND AN UNPLANNED INCISION OCCURRED WITH THE SURGEON USING OUTSIDE SCREWS TO GET THE COMPRESSION ACROSS THE ANKLE JOINT. THE CASE WAS COMPLETED USING ARTHREX 7.0 H
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 20577676 ↗InjuryOct 16, 2024Physical Resistance/Sticking; Physical Resistance/Sticking
ON (B)(6)2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-9091-02 HINDFOOT NAIL CABLE TENSIONER WOULD NOT RELEASE. THIS OCCURRED DURING A TTC FUSION ON (B)(6)2024 WHEN THE CABLES TENSIONER FOR THE TTC DUAL COMPRESSION NAIL WOULD NOT RELEASE AND SEEMED TO BE COLD WELDED. THEY WERE UNABLE TO USE THE DEVICE THE CORRECT WAY TO GAIN COMPRESSION. ONLY TWO SCREWS WERE IMPLANTED IN THE IM NAIL AND AUGMENTED WITH STAPLES TO ADD IN COMPRESSION. NAIL IS BEING REVISED AT A LATER DATE.
Patient problems: Insufficient Information; Pain
MAUDE report 20461235 ↗MalfunctionMar 28, 2024Material Deformation; Material Deformation
ON (B)(6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS (B)(4) THAT AN AR-9091-02 HINDFOOT NAIL CABLE TENSIONER SCREW IS STRIPPED. IT WAS HARD TO TIGHTEN AND LOOSEN. WAS DAMAGED WHEN PUTTING IT ON THE JIG. AS SOON AS IT WAS PLACED ON THE JIG IT POPPED OFF. THIS WAS DISCOVERED DURING A PROCEDURE WITH NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18997293 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page