Arthrex® AR-9091-02Recalls & reported problems

Arthrex, Inc. · Hindfoot Nail Cable Tensioner · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00888867335301), received between Mar 28, 2024 and Jul 25, 2025 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Physical Resistance/Sticking” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Physical Resistance/Sticking4
  • Defective Device1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Insufficient Information3
  • Pain1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 25, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    ON 06/30/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT AN AR-9091-02 HINDFOOT NAIL CABLE TENSIONER CABLE SETTING SCREWS IN THE TENSIONER HAVE BEEN REVERSE THREADED INTO THE JIG AND FREQUENTLY GETS STUCK. THIS HAPPENED DURING A CASE AND HAD NO ADVERSE EFFECTS ON THE PATIENT. ADDITIONAL INFORMATION WAS PROVIDED VIA PHONE, WHERE THE SALES REPRESENTATIVE STATED THIS EVENT OCCURRED DURING AN ANKLE FUSION PROCEDURE ON (B)(6) 2025, WHERE THEY WERE UNABLE TO COMPRESS THE NAIL. THIS RESULTED IN A DELAY OF ABOUT 15-20 MINUTES THAT DID NOT REQUIRE ADDITIONAL ANESTHESIA

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22608001 ↗
  • InjuryJan 6, 2025Defective Device; Physical Resistance/Sticking; Defective Device; Physical Resistance/Sticking

    ON 12/9/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN ARTHREX LOANER TRAY WAS RECEIVED A COUPLE OF DAYS BEFORE THE SURGERY TOOK PLACE ON (B)(6) 2024. IT WAS VERIFIED THAT THE 9091-02 HINDFOOT NAIL CABLE TENSIONER WAS IN A ZEROED-OUT POSITION BEFORE BEING STERILIZED. ABOUT THIRTY MINUTES BEFORE THE SURGERY, IT WAS VERIFIED AGAIN THAT THE 9091-02 HINDFOOT NAIL CABLE TENSIONER WAS IN A ZEROED-OUT POSITION. DURING THE PROCEDURE, WHEN THE TECHNICIAN WENT TO TIGHTEN DOWN THE SET SCREW WITH THE HEX DRIVER, THE DRIVER IMMEDIATELY BEGAN TORQUING WITHOUT ACTUALLY TIGHTENING THE SCREW ITSELF. THE TECHNICIAN TRIED TO LOOSEN THE SET SCREW, BUT IT WAS PROVING VERY DIFFICULT TO LEFTY LOOSEY THE SCREW EVEN AFTER THE FIRST COUPLE OF TURNS. IT WAS ATTEMPTED AGAIN TO TIGHTEN THE SCREW DOWN AGAIN NOW, AND IT IMMEDIATELY CLICKED DUE TO THE TORQUE LIMITING FEATURE ON THE HEX DRIVER LIKE LAST TIME. THE CABLE TENSIONER WAS ALSO ABLE TO BE PULLED FREELY ON THAT SIDE DESPITE THE TORQUE LIMITER CLICKING LIKE IT WAS FULLY SEATED. THE OTHER SIDE WAS ATTEMPTED TO BE TIGHTENED, BUT THE SAME ISSUE WAS HAPPENING. AFTER ABOUT FIVE MINUTES, THE SURGEON NOTICED THE SET SCREWS WERE VERY ELEVATED AN

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21079015 ↗
  • InjuryOct 31, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    ON 10/15/2024, A SALES REPRESENTATIVE REPORTED VIA SEMS-06686918 THAT AN AR-9091-02 HINDFOOT NAIL CABLE TENSIONER FAILED DURING A CASE. THE REP STATED THAT THE PATIENT WAS AFFECTED BY THE FAILURE. THEY COULD NOT USE THE NAIL'S TENSIONING COMPONENT AND HAD TO ADD ADDITIONAL HARDWARE TO COMPENSATE. THIS WAS DISCOVERED DURING A PROCEDURE. ADDITIONAL INFORMATION WAS RECEIVED ON 10/21/2024: THIS WAS DISCOVERED DURING A TTC NAIL PROCEDURE ON 10/14/2024. IT WAS STATED THAT THE FAILURE OF THE AR-9091-02 HINDFOOT NAIL CABLE TENSIONER WAS DUE TO THE SET SCREWS NOT MOVING UP AND DOWN WITHIN THE TENSIONER. THE SCREWS THEN ENDED UP SHEARING THE INNER WALL OF THE TENSIONER. THE SCREWS MAY HAVE ALSO BEEN STRIPPED. THE TENSIONER COULD NOT PULL THE NITINOL CORE OF THE NAIL TO GAIN COMPRESSION. THIS LEFT THE NAIL STATIC AND UNABLE TO USE THE SECOND CALCANEUS SCREW. THE CASE WAS COMPLETED SUCCESSFULLY BY USING ADDITIONAL FIXATION OUTSIDE OF THE NAIL. THE TECHNIQUE WAS CHANGED DUE TO NOT BEING ABLE TO USE THE COMPRESSION FEATURE OF THE NAIL, AND AN UNPLANNED INCISION OCCURRED WITH THE SURGEON USING OUTSIDE SCREWS TO GET THE COMPRESSION ACROSS THE ANKLE JOINT. THE CASE WAS COMPLETED USING ARTHREX 7.0 H

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20577676 ↗
  • InjuryOct 16, 2024Physical Resistance/Sticking; Physical Resistance/Sticking

    ON (B)(6)2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-9091-02 HINDFOOT NAIL CABLE TENSIONER WOULD NOT RELEASE. THIS OCCURRED DURING A TTC FUSION ON (B)(6)2024 WHEN THE CABLES TENSIONER FOR THE TTC DUAL COMPRESSION NAIL WOULD NOT RELEASE AND SEEMED TO BE COLD WELDED. THEY WERE UNABLE TO USE THE DEVICE THE CORRECT WAY TO GAIN COMPRESSION. ONLY TWO SCREWS WERE IMPLANTED IN THE IM NAIL AND AUGMENTED WITH STAPLES TO ADD IN COMPRESSION. NAIL IS BEING REVISED AT A LATER DATE.

    Patient problems: Insufficient Information; Pain

    MAUDE report 20461235 ↗
  • MalfunctionMar 28, 2024Material Deformation; Material Deformation

    ON (B)(6) 2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS (B)(4) THAT AN AR-9091-02 HINDFOOT NAIL CABLE TENSIONER SCREW IS STRIPPED. IT WAS HARD TO TIGHTEN AND LOOSEN. WAS DAMAGED WHEN PUTTING IT ON THE JIG. AS SOON AS IT WAS PLACED ON THE JIG IT POPPED OFF. THIS WAS DISCOVERED DURING A PROCEDURE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18997293 ↗

Other Rod, Fixation, Intramedullary And Accessories devices