Skip to content

Arthrex® AR-9107-02-25RRecalls & reported problems

Arthrex, Inc. · AUGMENTED VAULTLOCK, MEDIUM 25R · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867350717), received between Jan 12, 2026 and Jan 12, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 12, 2026Break

    ON (B)(60 2025, A DISTRIBUTOR REPORTED VIA EMAIL THAT AN AR-9107-02-25R UNIVERS VAULTLOCK AUGMENTED GLENOID, MEDIUM 25R HAD ITS CENTRAL PEG BREAK OFF THE IMPLANT PRIOR TO IMPLANTATION (SEE ATTACHED PHOTO). THE ISSUE WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM. ADDITIONAL INFORMATION WAS RECEIVED ON 24-DEC-2025: DURING A TOTAL SHOULDER ARTHROPLASTY UTILIZING AN AUGMENTED VAULTLOCK GLENOID AND ECLIPSE SYSTEM, THE CENTRAL PEG OF AN AR-9107-02-25R UNIVERS VAULTLOCK AUGMENTED GLENOID, MEDIUM 25R BROKE WHILE APPLYING GRAFT WITH THE GRAFT COMPRESSION TOOL PRIOR TO IMPLANTATION. ALL COMPONENTS WERE RETRIEVED BEFORE USE. A REPLACEMENT AR-9107-02-15R UNIVERS VAULTLOCK AUGMENTED GLENOID, MEDIUM 15R WAS IMPLANTED INSTEAD. THE EVENT RESULTED IN AN APPROXIMATE 5¿10 MINUTE DELAY, WITH NO ADDITIONAL ANESTHESIA ADMINISTERED. BONE QUALITY WAS ASSESSED AS NORMAL, AND NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24036219 ↗

Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices

All PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page