Arthrex® AR-9094-090Recalls & reported problems

Arthrex, Inc. · Telescoping lag screw, 10.5 mm x 90 mm · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888867409996), received between Feb 26, 2025 and Mar 30, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Physical Resistance/Sticking” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Physical Resistance/Sticking3
  • Break1

Patient problems

  • Insufficient Information2
  • Failure of Implant1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 30, 2026Physical Resistance/Sticking

    IT WAS REPORTED THAT DURING THE NAILING OF THE PROXIMAL FEMUR ON (B)(6) 2026, THE TELESCOPING LAG SCREW WAS INSERTED INTO THE BONE AT THE WRONG ANGLE. DUE TO THIS FAILURE THE SURGEON ATTEMPTED TO REMOVE THE SCREW, HOWEVER DUE TO THE INCORRECT INSERTION THE SCREW BECAME JAMMED. THE SURGEON DECIDED THAT THE FIXATION SHOULD BE SUFFICIENT AND THEREFORE THE SCREW WAS LEFT IN PLACE. IT WAS FURTHER REPORTED THAT DURING THE ATTEMPT OF REMOVING THE SCREW, THE INSTRUMENTS USED WERE BENT AS WELL. NOT LONG AFTER THE INITIAL SURGERY, IT WAS DISCOVERED THAT THE IMPLANTED DEVICE HAD FAILED AND THE NAIL HAD DETACHED FROM THE BONE. FOR THIS REASON, A REVISION SURGERY WAS PERFORMED ON (B)(6) 2026, DURING WHICH ALL PARTS WERE REPLACED WITH NEW DEVICES WITH THE SAME PART NUMBERS. EXCEPT THE SURGEON DECIDED TO USE A NAIL WITH A LONGER SHAFT TO ENSURE PROPER FIXATION. THE REVISION SURGERY WAS PERFORMED SUCCESSFUL, AND NO FURTHER ACTION IS PLANNED.

    Patient problems: Failure of Implant

    MAUDE report 24711887 ↗
  • MalfunctionMar 21, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    ON 3/4/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-9094-090 TELESCOPING LAG SCREW WOULD NOT COLLAPSE AFTER REMOVING THE TELESCOPING ACTUATOR. THE SCREW WAS REMOVED AND REPLACED WITH A NEW SCREW. THIS WAS DISCOVERED DURING A PROCEDURE ON (B)(6) 2025. NO FURTHER. INFORMATION WAS REPORTED. ADDITIONAL INFORMATION WAS REQUESTED ON 3/12/2025.

    Patient problems: Insufficient Information

    MAUDE report 21663435 ↗
  • MalfunctionFeb 26, 2025Break; Physical Resistance/Sticking; Break; Physical Resistance/Sticking

    ON 01/29/2025, AN FDA MEDWATCH NOTIFICATION WAS RECEIVED VIA EMAIL. A FACILITY REPRESENTATIVE REPORTED THAT AN 9094-090 TELESCOPING LAG SCREW FELL APART DURING INSERTION. A C-ARM WAS BEING USED DURING THE PROCEDURE; HOWEVER, NO FRAGMENTS WERE SEEN ON THE X-RAY. THIS WAS DISCOVERED DURING A PROCEDURE IN (B)(6) 2025. NO FURTHER INFORMATION WAS REPORTED. ADDITIONAL INFORMATION RECEIVED ON 02/06/2025: THIS WAS DISCOVERED DURING A HIP TROCHANTERIC NAIL PROCEDURE ON (B)(6) 2025. A PIECE OF THE LAG SCREW BROKE DURING INSERTION. THE SCREW WAS REMOVED AND REPLACED WITH A NEW ONE TO COMPLETE THE CASE. AN X-RAY CONFIRMED THAT NO FRAGMENTS WERE LEFT BEHIND.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21468672 ↗

Other Rod, Fixation, Intramedullary And Accessories devices

All Rod, Fixation, Intramedullary And Accessories devices →