Skip to content

Arthrex® AR-9094-10-2030Recalls & reported problems

Arthrex, Inc. · Trochanteric nail, 10mm x 20cm x 130° · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867411548), received between Jun 7, 2024 and Jun 7, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Use of Device Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 7, 2024Use of Device Problem; Use of Device Problem

    ON 05/09/2024, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN 1192-100 TELESCOPING LAG SCREW, A 1099-110 TELESCOPING LAG SCREW, AND A 1099-100 TELESCOPING LAG SCREW WOULD NOT ADVANCE INTO THE FEMORAL HEAD OF AN AR-9094-10-2030 TROCHANTERIC NAIL. THE TELESCOPING LAG SCREWS WERE USED WITH A NEW TROCHANTERIC NAIL; HOWEVER, THE ISSUE PERSISTED, AND THE TELESCOPING LAG SCREWS WOULD STILL NOT ADVANCE. THE CASE WAS COMPLETED USING A NEW TROCHANTERIC NAIL AND A NEW 1099-110 TELESCOPING LAG SCREW. THE CASE WAS DELAYED ABOUT FIFTEEN MINUTES. THIS WAS DISCOVERED DURING A LEFT HIP IM NAIL PROCEDURE ON (B)(6) 2024.

    Patient problems: Insufficient Information

    MAUDE report 19487721 ↗

Other Rod, Fixation, Intramedullary And Accessories devices

All Rod, Fixation, Intramedullary And Accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page