Arthrex® AR-9094ES-11-3930LRecalls & reported problems
Arthrex, Inc. · ES troch nail, Lt, 11mm x 39cm x 130° · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867412750), received between Oct 22, 2025 and Oct 22, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Use of Device Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 22, 2025Use of Device Problem; Use of Device Problem
ON 10/17/2025, A SALES REPRESENTATIVE REPORTED VIA (B)(4) THAT A 1268-100 RADIOLUCENT TARGETING ARM, 130 DEGREE TROCH NAIL AND AN AR-9094ES-11-3930L ES TROCHANTERIC NAIL COULD NOT BE USED AS INTENDED. THE ISSUE OCCURRED DURING INSERTION OF THE ES SCREW, WHERE THE SCREW TRAJECTORY WAS INACCURATE, CAUSING IT TO BIND WITHIN THE NAIL. THE SCREW CROSS-THREADED, AND THE SURGEON WAS UNABLE TO FULLY ADVANCE IT THROUGH THE NAIL. THE CASE WAS COMPLETED WITHOUT DISTAL LOCKING OF THE NAIL. ATTEMPTS TO ALIGN THE DRILL ADDED A DELAY OF APPROXIMATELY FIVE TO TEN MINUTES, RESULTING IN AN ADDITIONAL FIVE TO TEN MINUTES OF ANESTHESIA ADMINISTRATION. THE SAME ORIGINAL PART NUMBERS WERE USED TO COMPLETE THE CASE. NO FURTHER INFORMATION IS AVAILABLE. NO PHOTOS WERE TAKEN FOR DOCUMENTATION, AND NO ADVERSE EFFECTS WERE REPORTED. THIS ISSUE WAS IDENTIFIED DURING A HIP FRACTURE PROCEDURE ON (B)(6) 2025, WITH NO REPORTED PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23356435 ↗
Other Rod, Fixation, Intramedullary And Accessories devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page