Skip to content

Arthrex® AR-3373-4202Recalls & reported problems

Arthrex, Inc. · SHEATH,HI-FLOW,TAP/FEN,2STOPCOCKF/UWV · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00888867431270), received between Jan 28, 2025 and May 27, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 3 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Detachment of Device or Device Component2
  • Device Damaged by Another Device1
  • Material Frayed1
  • Sharp Edges1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Insufficient Information1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 27, 2026Break

    ON 28-APR-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-3373-4202 SHEATH'S O-RING BROKE FREE AND WENT INTO THE PATIENT WITH WATER FLOW. NOTICED DURING A CASE AND COMPLETED WITH NO PATIENT EFFECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25301005 ↗
  • MalfunctionMay 27, 2026Detachment of Device or Device Component

    ON 28-APR-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-3373-4202 SHEATH'S O-RING CAME OUT AND ENDED UP INSIDE THE PATIENT'S SHOULDER. THIS WAS DISCOVERED DURING A PROCEDURE WITH NO PATIENT HARM REPORTED. THE O-RING WAS ABLE TO BE SUCCESSFULLY RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25300847 ↗
  • MalfunctionMay 1, 2026Material Frayed

    ON 6-APR-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-3373-4202 SHEATH IS FRAYED AT THE DISTAL TIP. NOTICED DURING A CASE WITH NO PATIENT EFFECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25055627 ↗
  • MalfunctionApr 30, 2026Break

    ON 13-APR-2026, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-3373-4202 SHEATH IS DAMAGED, HAS A PIECE MISSING. NOTICED DURING A CASE WITH NO PATIENT EFFECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25042804 ↗
  • MalfunctionNov 24, 2025Device Damaged by Another Device; Device Damaged by Another Device

    IT WAS REPORTED THAT DURING A SURGERY THE PATIENT EXPERIENCED AN INVOLUNTARY ARM MOVEMENT WHICH CAUSED THE ¿FISH EYE¿ END OF THE PANOSCOPE TO COME INTO CONTACT WITH THE SHEATH. THIS RESULTED IN DAMAGE TO BOTH THE SCOPE AND THE SHEATH. PER COMPLAINT REPORTER THERE WAS NO HARM FOR PATIENT, OPERATOR OR THIRD PARTY. NO FURTHER INFORMATION RECEIVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23631306 ↗
  • MalfunctionJul 14, 2025Sharp Edges; Sharp Edges

    ON 06/17/2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA SEMS-(B)(4) THAT AN AR-3373-4202 SHEATH IS SHARP AT THE TIP (FENESTRATED PART). WHEN GOING IN AND OUT OF THE PATIENT, IT IS CAUSING ADDITIONAL DAMAGE TO TISSUE AND CARTILAGE. DISCOVERED DURING A CASE AND DEVICE SWAPPED.

    Patient problems: Insufficient Information

    MAUDE report 22494282 ↗
  • MalfunctionJan 28, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    ON 01/02/2025 IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-3373-4202 SHEATH STOPCOCK FELL OFF. THIS WAS DISCOVERED DURING THE CASE WITH NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21255814 ↗

Other Accessories, arthroscopic devices

All Accessories, arthroscopic devices →