Skip to content

Arthrex® AR-S4000K-SRecalls & reported problems

Arthrex, Inc. · Spine Access Kit, Disposable · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888867465374), received between Mar 6, 2025 and Mar 6, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 6, 2025Break; Break

    ON (B)(6)2025, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA (B)(4) THAT AN AR-S4000K-S SPINE ACCESS KIT NITINOL K-WIRE GOT KINKED AND EVENTUALLY BROKE OFF IN THE PATIENT, ABOUT 3 CM FROM THE DISTAL END WHILE DOING A TRANSFORAMINAL APPROACH. THIS OCCURRED DURING USE IN A L5/S1 TRANSFORAMINAL DISCECTOMY CASE WITH NO PATIENT EFFECT. THE PROCEDURE WAS COMPLETED ENDOSCOPICALLY. THE BROKEN PIECE WAS NOT RETRIEVED FROM THE PATIENT. ADDITIONAL INFORMATION REQUESTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 21536688 ↗