Skip to content

DJO SURGICAL 804-03-008Recalls & reported problems

Encore Medical, L.P. · GUIDE, REAMER, BONE TAP, 6.5MM, RSP, SHOULDER · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
17
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 00888912147842), received between Feb 9, 2017 and Oct 11, 2024 — classified as 3 injuries, 14 malfunctions.

The most frequently reported specific device problem is “Break” (13 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break13
  • No Apparent Adverse Event2
  • Detachment of Device or Device Component1
  • Device Handling Problem1
  • Material Integrity Problem1

Patient problems

  • Insufficient Information5
  • No Known Impact Or Consequence To Patient5
  • No Clinical Signs, Symptoms or Conditions2
  • Device Embedded In Tissue or Plaque1
  • Foreign Body In Patient1
  • No Information1
  • No Patient Involvement1
  • Patient Problem/Medical Problem1

Recent adverse event reports

The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 11, 2024No Apparent Adverse Event; No Apparent Adverse Event

    INSTRUMENT FAILURE - WHEN DOCTOR REAMING THE TAP BROKE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20434666 ↗
  • InjuryApr 6, 2024No Apparent Adverse Event; No Apparent Adverse Event

    IT WAS REPORTED THERE WAS A 30 MINUTE DELAY IN SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 19056507 ↗
  • InjuryMay 19, 2023Material Integrity Problem; Material Integrity Problem

    REPORTED INCIDENT - 30 MINUTE DELAY IN SURGERY, PART LEFT IN PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 16969435 ↗
  • MalfunctionDec 14, 2022Device Handling Problem; Device Handling Problem

    INSTRUMENT FAILURE: THE TAP BROKE IN HALF INSIDE OF THE BROWN REAMER. PART OF THE BROKEN TAP IS STUCK INSIDE OF THE BROWN REAMER, LEAVING BOTH INSTRUMENTS BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15980335 ↗
  • MalfunctionJan 12, 2022Break

    INSTRUMENT FAILURE - DUE TO BROKEN INSTRUMENT.

    Patient problems: Insufficient Information

    MAUDE report 13237925 ↗
  • MalfunctionSep 1, 2021Break

    INSTRUMENT FAILURE - TAP BROKE DURING THE CASE IN THE GLENOID.

    Patient problems: Insufficient Information

    MAUDE report 12410945 ↗
  • MalfunctionApr 8, 2021Break

    INSTRUMENT FAILURE - DUE TO DURING THE TAPPING PROCESS, THE TAP SNAPPED OFF INSIDE THE GLENOID. DR. REDIRECTED THE DRILL AND PLACED THE BASEPLATE INFERIOR OF THE BROKEN TAP.

    Patient problems: Insufficient Information

    MAUDE report 11638426 ↗
  • MalfunctionJul 23, 2020Break

    INSTRUMENT FAILURE - SURGEON WAS IN THE PROCESS OF REAMING THE GLENOID AND THE TAP SNAPPED OFF AT THE THREAD. SURGEON WAS ABLE TO REDIRECT AN ALTERNATE TAP AND COMPLETE PROCEDURE. THE REMAINING TAP WAS NOT RETRIEVABLE AND THEREFORE LEFT BEHIND. THE REMAINDER OF THE TAP WILL BE RETURNED TO DJO.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10312119 ↗
  • MalfunctionFeb 19, 2020Break

    INSTRUMENT FAILURE - THE SURGEON WAS REAMING AND THE TAP SNAPPED WHERE THE THREADS BEGIN. NO PIECE WAS LEFT IN THE PATIENT.

    Patient problems: No Patient Involvement

    MAUDE report 9729343 ↗
  • MalfunctionJun 24, 2019Break

    INSTRUMENT FAILURE - THE 6. 5 MM TAP SNAPPED OFF IN GLENOID.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8728698 ↗
  • MalfunctionJun 24, 2019Break

    INSTRUMENT FAILURE - TAP BROKE OFF DUE TO SCLEROTIC BONE WHILE REAMING.

    Patient problems: Patient Problem/Medical Problem

    MAUDE report 8727900 ↗
  • MalfunctionJun 21, 2019Break

    PROBLEM - BROKE DURING SURGERY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8725113 ↗
  • MalfunctionMar 28, 2019Break

    INSTRUMENT FAILURE - THE REVERSE TAP BROKE AT THE DEPTH ETCHING WHILE REAMING. IT WAS BROKEN BELOW THE BONE SO IT WAS LEFT IN AND A NEW HOLE FOR THE BASEPLATE DRILLED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8461571 ↗
  • MalfunctionJan 7, 2019Break; Detachment of Device or Device Component

    INSTRUMENT FAILURE - WHILE REAMING THE GLENOID, THE TAB BROKE. PART WAS LEFT IN PATIENT, AND PART DISCARDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 8222865 ↗
  • MalfunctionJul 23, 2018Break

    INSTRUMENT FAILURE - THE SURGEON WAS REAMING THE GLENOID OVER THE TAP, WHEN THE TAP BROKE, LEAVING THE SCREW THREAD PORTION STUCK IN PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 7712244 ↗
  • MalfunctionNov 7, 2017Break

    INSTRUMENT FAILURE - WHILE THE SURGEON WAS REAMING OVER THE TAP, THE TAP BROKE IN THE GLENOID.

    Patient problems: No Information

    MAUDE report 7005993 ↗
  • MalfunctionFeb 9, 2017Break

    INSTRUMENT FAILURE - THE 6.5 TAP BROKE OFF IN THE PATIENT'S GLENOID HEAD, BUT THE SURGEON WAS ABLE TO GET IT OUT BECAUSE IT WAS FLUSH WITH THE GLENOID. THE SURGEON TOOK A TREPHINE AND DRILLED OVER IT WHICH THEN GRABBED THE GLENOID AND IT CAME OUT. THE SURGEON WAS THEN ABLE TO PUT IMPLANT IN WITHOUT ANY ISSUE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6316766 ↗

Other Guide, surgical, instrument devices

All Guide, surgical, instrument devices →

Sources and updates