Skip to content

DJO SURGICAL 351-04-106Recalls & reported problems

Encore Medical, L.P. · DJO EMPOWR KNEETM, FIN BP MINUS, NP 6R · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00888912168137), received between Nov 23, 2020 and Mar 5, 2024 — classified as 3 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth2
  • Adverse Event Without Identified Device or Use Problem1
  • Appropriate Device Problem Term/Code Not Available1
  • Therapeutic or Diagnostic Output Failure1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Insufficient Information1
  • Joint Laxity1
  • Loss of Range of Motion1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY - PATIENT KNEE WAS VERY STIFF AND WE REMOVED THE TIBIA IN ORDER TO CUT MORE BONE AWAY TO LOOSEN IT UP.

    Patient problems: Insufficient Information; Loss of Range of Motion

    MAUDE report 18838670 ↗
  • InjuryDec 21, 2021Appropriate Device Problem Term/Code Not Available; Loosening of Implant Not Related to Bone-Ingrowth

    REVISION SURGERY:DUE TO TIBIAL LOOSENING.

    Patient problems: Joint Laxity

    MAUDE report 13052694 ↗
  • InjuryNov 23, 2020Therapeutic or Diagnostic Output Failure; Loosening of Implant Not Related to Bone-Ingrowth

    REVISION SURGERY - REVISED A LOOSE SIZE 6 RIGHT TIBIA BASEPLATE THAT LOOSENED IN THE CEMENT MANTLE. BELIEVED TO BE A CEMENT MANTLE FAILURE AND NOT A COMPONENT FAILURE. DONE ROUGHLY 2 MONTHS AGO. PUT IN A UNIVERSAL BASEPLATE WITH STEM AND NEW 3D POLY.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 10884820 ↗

Other Guide, Surgical, Instrument devices

All Guide, Surgical, Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page