Skip to content

DJO SURGICAL 804-06-056Recalls & reported problems

Encore Medical, L.P. · Encore Reverse Shoulder Humeral Trial/Implant Inserter, Std · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00888912171564), received between Apr 6, 2017 and Apr 25, 2025 — classified as 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Detachment Of Device Component1
  • Detachment of Device or Device Component1
  • Device Operates Differently Than Expected1
  • Mechanical Problem1
  • Physical Property Issue1

Patient problems

  • Insufficient Information3
  • No Known Impact Or Consequence To Patient3
  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1
  • No Information1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 25, 2025Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component

    INSTRUMENT FAILURE - THE KNOB ON THE STEM INSERTER ALSO BROKE OFF INTO THE STEM ONCE IT WAS IMPACTED. BROKEN FRAGMENTED PIECES LEFT IN THE PATIENT.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21915079 ↗
  • MalfunctionFeb 24, 2022Break

    INSTRUMENT FAILURE: THE SCREW ON THE KNOB OF THE INSERTER SHEARED OFF INSIDE OF THE IMPLANTED STEM. THIS PREVENTED THE 8MM SPACER SCREW FROM BEING INSERTED INTO THE STEM. THE STEM HAD TO BE EXPLANTED AND A NEW STEM IMPLANTED TO ALLOW USE OF THE 8MM SPACER.

    Patient problems: Insufficient Information

    MAUDE report 13590900 ↗
  • MalfunctionAug 23, 2021Break

    INSTRUMENT FAILURE - WHEN IMPACTING THE FINAL BROACH IN TO THE CANAL THE KNOB ON THE INSERTER BROKE AT THE "DEAD ZONE" BETWEEN THREAD PITCHES. THIS IS THE SECOND KNOB THAT HAS BROKEN DURING SURGERY.

    Patient problems: Insufficient Information

    MAUDE report 12356062 ↗
  • MalfunctionJun 2, 2021Break

    INSTRUMENT FAILURE - QUICK COUPLING T-HANDLE BROKE DURING REMOVAL OF 6.5MM BONE TAP FROM GLENOID. PATIENT HAD EXCEPTIONALLY HARD BONE AND CAUSED THE CONNECTION BETWEEN THE HANDLE" AND THE SHAFT BROKE, CAUSING "HANDLE" TO ROTATE FREELY AROUND CONNECTION TO SHAFT. THREADED IMPACTION HANDLE KNOB BROKE BETWEEN PROXIMAL AND DISTAL SCREWS DURING IMPLANT INSERTION. DISTAL PORTION OF THREADS ARE STUCK IN FINAL IMPLANT STEM. BREAK HAD NO EFFECT ON MATING BETWEEN POLY INSERT AND STEM.

    Patient problems: Insufficient Information

    MAUDE report 11919120 ↗
  • MalfunctionJan 13, 2020Break

    INSTRUMENT FAILURE - DUE TO STEM INSERTER TIP KNOB SHEARING OFF DURING THE USE OF SURGERY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9582926 ↗
  • MalfunctionApr 23, 2018Device Operates Differently Than Expected; Physical Property Issue

    INSTRUMENT FAILURE - DUE TO THE SCREW ON INSERTER SHEARING OFF UPON IMPACTION. REMNANT OF SCREW WAS LEFT IN HUMERAL STEM WITH NO ADVERSE EFFECTS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7451188 ↗
  • MalfunctionDec 27, 2017Break

    INSTRUMENT FAILURE - THE SURGEON WAS PERFORMING A REVERSE TOTAL SHOULDER PROCEDURE WHEN THE SCREW IN THE ALTIVATE IMPACTOR HANDLE BROKE UPON IMPACTION. THE CASE WAS WITHOUT INCIDENT UNTIL THE SURGEON IMPACTED THE STEM INTO THE HUMERUS. AS HE WAS HITTING THE IMPACT PLATE OF THE HANDLE, THE RETENTION SCREW LOOSENED, THUSLY ALLOWING ADDITIONAL FORCES TO ACT UPON THE SCREW. THE BALANCE OF THE SCREW WAS LEFT IN THE IMPLANT (530-12-108). THE SURGEON DICTATED THE INCIDENT IN HIS NOTES.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7147577 ↗
  • MalfunctionApr 6, 2017Break; Detachment Of Device Component

    PRIMARY SURGERY - THE SURGEON WAS INSERTING THE STEM AND THE THREADS ON THE STEM INSERTER SHEARED OFF THE STEM IMPLANT. THE IMPLANT WAS WASTED FROM THIS AND A DIFFERENT STEM IMPLANT WAS USED.

    Patient problems: No Information

    MAUDE report 6464884 ↗

Other Guide, surgical, instrument devices

All Guide, surgical, instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page