K-Systems K22074Recalls & reported problems
Coopersurgical, Inc. · G85 Mini Incubator · Class 2
All models of K-Systems (6)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00888937025057), received between May 24, 2023 and May 24, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Rupture” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Recalls naming this device
- Nov 15, 2024Class IIOpen, Classified
The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.
Z-0688-2025 · Root cause: Process change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 24, 2023Material Rupture; Material Rupture
IN AN ATTEMPT TO GET THE GAS FLOW RUNNING, PURGE WAS PRESSED, AND THE BOTTLE LID BLEW OFF. THE REGULATOR WAS AT OVER 1 BAR, WHEREAS THE MANUAL STATES IT SHOULD BE LESS THAN 0.5 BAR. THEN, WHILE REMOVING THE INLET TUBE, THE WELDED BOTTLE JOIN BROKE OFF. 1216677-2023-00086 G85 MINI INCUBATOR K22074 (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16991858 ↗
FDA clearance and classification
The FDA regulates product code PUB (accessory, assisted reproduction, exempt) under 21 CFR 884.6120 as a Class II device. 21 CFR 884.6120 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
FDA definition of this device type: This product code is the class II exempt counterpart of MQG, and was exempted, subject to the limitations of exemption, under the procedures established by the 21st Century Cures Act. This device type does not require premarket notification (510(k)) to market if it does not exceed the general limitations of exemption found in 21 CFR 884.9 and the specific limitations of exemption as stated in the Physical State field of this product code description.
Common questions about the K-Systems K22074
Has the K-Systems K22074 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0688-2025 (Class II), began on Nov 15, 2024: The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the K-Systems K22074?
The FDA's MAUDE database lists 1 adverse event report for this device, received on May 24, 2023. That report lists device problem “Material Rupture” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the K-Systems K22074?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the K-Systems K22074 FDA approved?
Devices of this type (product code PUB) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Coopersurgical, Inc.
Who makes the K-Systems K22074, and is it still sold?
The FDA's device database lists Coopersurgical, Inc. as the company responsible for it (brand K-Systems, model K22074). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Accessory, Assisted Reproduction, Exempt devices
- K-Systems K23054Coopersurgical, Inc. · Accessory, Assisted Reproduction, Exempt1 recall0 reports
- K-Systems K23055Coopersurgical, Inc. · Accessory, Assisted Reproduction, Exempt1 recall0 reports
- K-Systems K24012Coopersurgical, Inc. · Accessory, Assisted Reproduction, Exempt1 recall0 reports
- K-Systems K22065Coopersurgical, Inc. · Accessory, Assisted Reproduction, Exempt1 recall0 reports
- K-Systems 6.131.812.121Coopersurgical, Inc. · Accessory, Assisted Reproduction, Exempt1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice