K-Systems recallZ-0688-2025 · CooperSurgical, Inc. · initiated Nov 15, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from CooperSurgical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.
FDA root cause category: Process change control
What the company is doing
On November 15, 2024 URGENT MEDICAL DEVICE RECALL emails were sent to customers. On November 20, 2024 URGENT MEDICAL DEVICE RECALL letters were sent to customers. Actions that should be taken by the customer/user/distributor in order to prevent risks for patients or users: G85 Mini Incubators with serial numbers listed in the Appendix are impacted and require repair. 1. Inspect your inventory, identify, and discontinue use. 2. Complete the attached Urgent Medical Device Recall Response Form and return the form to CooperSurgical as indicated at the top of the form. Note: Please complete and return the form even if you do not have any affected product in your inventory so that we may document receipt of this Urgent Medical Device Recall. 3. Contact your local Service Representative to arrange a repair at no cost to you. T45, T47, R65, and G73 with serial numbers listed in the Appendix may be impacted and require inspection to determine the need for repair. 1. Inspect your inventory, identify, and discontinue use. 2. Complete the attached Urgent Medical Device Recall Response Form and return the form to CooperSurgical as indicated at the top of the form. Note: Please complete and return the form even if you do not have any affected product in your inventory so that we may document receipt of this Urgent Medical Device Recall. 3. Contact your local Service Representative to arrange an inspection of the affected product. If your product is affected, the product will be repaired at no cost to you. 6. Actions planned by CooperSurgical to correct the problem CooperSurgical is currently investigating this issue and determined a field action notice to customers is needed to inspect and repair all affected products in the field. CooperSurgical is committed to high-quality, safe, and effective products. This corrective action has been initiated to ensure this failure mode does not reoccur and future potential patient harm can be avoided. If you need any further
Recall details
- Recall number
- Z-0688-2025
- Status
- Open, Classified
- Initiated by the company
- Nov 15, 2024
- Posted by the FDA
- Dec 18, 2024
- Recalling firm
- CooperSurgical, Inc.
- Quantity
- 37 units
- Product code
- PUB
- Distribution
- States: NC, NJ, TX, VA, IL, NY, CA, PA Please note: The countries below, excluding Iraq, all consider these devices to be "lab equipment", not medical devices and are not subject to classification or recall guidelines. Countries: AT, AU, BE, BH, BR, CA, CH, CN, CZ, DE, EG, ES, ET, FR, GB, GH, HU, ID, IE, IL, IN, IQ, IT, JO, JP, KR, KW, LB, MA, MK, MT, NL, NO, PE, PH, PK, PT, QA, RO, RS, SA, SD, SE, SO,TR, TW, TZ, US, ZA
Product description
Brand Name: K-Systems Product Name: G85 Mini Incubator Model/Catalog Number: K22074 Software Version: N/A Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes and embryos), at closely regulated and monitored temperatures, during assisted reproduction examination and processing procedures. It may support a more comprehensive growth environment [e.g., temperature, carbon dioxide (CO2) and oxygen (O2) concentration]; it typically has adjustment and monitoring controls. Component: No
Code information (models, lots, serial numbers)
Version or Model: K22074; Primary DI Number: 00888937025057; Serial Numbers: 2244KH01 , 2244KH02 , 2244KH03 , 2244KH04 , 2244KH05 , 2244KH06 , 2244KH07 , 2244KH08 , 2244KH09 , 2244KH10 , 2313KH01 , 2313KH02 , 2313KH03 , 2313KH04 , 2313KH05 , 2313KH06 , 2313KH07 , 2313KH08 , 2313KH09 , 2313KH10 , 2320KH21 , 2320KH22 , 2320KH23 , 2320KH24 , 2320KH25 , 2320KH26 , 2320KH27 , 2320KH28 , 2320KH29 , 2320KH30 , 2321KH16 , 2321KH17 , 2321KH18 , 2321KH19 , 2321KH20.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- K-Systems K22074Coopersurgical, Inc. · Accessory, Assisted Reproduction, Exempt1 recall1 report
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 95741).
- Z-0687-2025 · Open, ClassifiedBrand Name: K-Systems Product Name: G73 Dry Bath Incubator Model/Catalog Number: K22065 Product Description: An electrically-powered device intended to provide suitable warming and…
- Z-0689-2025 · Open, ClassifiedBrand Name: K-Systems Product Name: T47 Warming Plate Model/Catalog Number: K23054 Software Version: N/A Product Description: An electrically-powered device intended to provide sui…
- Z-0690-2025 · Open, ClassifiedBrand Name: K-Systems Product Name: T47 Warming Plate Model/Catalog Number: K23055 Software Version: N/A Product Description: An electrically-powered device intended to provide sui…
- Z-0691-2025 · Open, ClassifiedBrand Name: K-Systems Product Name: R65 Trolley Model/Catalog Number: K24012 Software Version: N/A Product Description: An electrically-powered device intended to provide suitable …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗