Vitality® 07.02054.001Recalls & reported problems
Zimmer Spine, Inc. · Class 2
All models of Vitality (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00889024009806), received between Nov 23, 2021 and Nov 14, 2025 — classified as 1 injury, 12 malfunctions.
The most frequently reported specific device problem is “Fracture” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture6
- Material Deformation4
- Device-Device Incompatibility2
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- Insufficient Information3
- Foreign Body In Patient1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 14, 2025Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT TWO VITALITY STANDARD REDUCTION SCREWDRIVERS WOULD NOT LOAD A SCREW PROPERLY. ANOTHER DRIVER WAS USED AND THERE WAS NO PATIENT IMPACT. UPON MANUFACTURER INVESTIGATION, IT WAS DISCOVERED THAT ONE OF THE SCREWDRIVERS WAS DEFORMED AND FRACTURED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23556935 ↗MalfunctionSep 11, 2025Fracture; Fracture
IT WAS REPORTED THAT THE TIP OF VITALITY SCREW DRIVER BROKE OFF IN A SCREW INTRA-OPERATIVELY. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23033600 ↗MalfunctionMay 7, 2025Material Twisted/Bent; Material Twisted/Bent
A DISTRIBUTOR REPORTED THAT A VITALITY DRIVER WAS FOUND TO HAVE A TWISTED TIP DURING SPD. THERE WAS NO REPORTED PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21979302 ↗InjuryMar 13, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGERY, THE TIP OF A FINAL VITAL DRIVER FRACTURED. THIS RESULTED IN A 20 MINUTE DELAY AND THEY WERE UNABLE TO RETRIEVE THE FRACTURED TIP. THE SURGERY WAS COMPLETED USING OTHER INSTRUMENTS AND THERE WAS NO PATIENT IMPACT BEYOND THE TIP RETENTION.
Patient problems: Foreign Body In Patient
MAUDE report 21595482 ↗MalfunctionAug 8, 2024Fracture; Fracture
IT WAS REPORTED THAT A VITALITY SCREWDRIVER TIP WAS DISCOVERED TO BE FRACTURED WHEN IT WAS RETURNED TO A WAREHOUSE. IT IS UNKNOWN WHEN THE DAMAGE OCCURRED, BUT THERE WAS NO REPORTED PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19939257 ↗MalfunctionJun 12, 2024Fracture; Fracture
IT WAS REPORTED THE BLUE PLASTIC STOP INSIDE A DRIVER IS BROKEN INTO 2 PIECES PREVENTING THE SLEEVE FROM SLIDING ALL THE WAY DOWN TO ENGAGE WITH THE TULIP OF THE SCREW. IT WAS FOUND DURING A ROUTINE INSPECTION SO THE PATIENT AND SURGICAL INFORMATION IS UNKNOWN. DEVICE INSPECTION BY THE MANUFACTURER FOUND THAT THE TIP OF THE DRIVER IS DEFORMED.
Patient problems: Insufficient Information
MAUDE report 19518814 ↗MalfunctionJan 25, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A SURGERY, THREE VITALITY DRIVERS FRACTURED WHEN TRYING TO REMOVE A STUCK ILIAC SCREW. THE SCREW WAS ULTIMATELY ABLE TO BE REMOVED AND REPLACED WITH A SHORTER SCREW. THIS RESULTED IN A 20 MINUTE DELAY TO THE PROCEDURE BUT NO OTHER PATIENT IMPACT. THIS IS REPORT TWO OF FOUR FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18584899 ↗MalfunctionMar 22, 2023Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT THE TIP OF A VITALITY SCREWDRIVER WAS DIFFICULT TO CONNECT TO A MATING SCREW WHEN PUTTING THE KIT BACK TOGETHER AFTER SURGERY. THERE WERE NO IMPACTS DURING THE SURGERY. HOWEVER, DEVICE EVALUATION BY THE MANUFACTURER FOUND THE TIP WAS DEFORMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16592344 ↗MalfunctionFeb 13, 2023Material Deformation; Material Deformation
IT WAS REPORTED THAT TWO VITALITY CONNECTORS BROKE, THE TIP OF ONE VITALITY T20 FINAL DRIVER BROKE, THE TIP OF A VITALITY T27 DRIVER TWISTED, AND THE TIPS OF TWO VITAL NAVIGATION T27 DRIVERS TWISTED INTRA-OPERATIVELY. OTHER CONNECTORS AND DRIVERS WERE USED TO COMPLETE THE PROCEDURE. THERE WAS A DELAY TO THE PROCEDURE THAT DID NOT LEAD TO PATIENT IMPACTS. THIS IS REPORT THREE OF SIX FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16362869 ↗MalfunctionOct 7, 2022Fracture; Fracture
IT WAS REPORTED THAT THE TIP OF A VITALITY SCREWDRIVER WAS FOUND FRACTURED WHEN PUTTING THE KIT BACK TOGETHER AFTER SURGERY. THERE WERE NO IMPACTS DURING THE SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15568141 ↗MalfunctionNov 29, 2021Material Deformation; Material Deformation
IT WAS REPORTED THE TIPS OF TWO SCREW DRIVERS WERE FOUND TO BE STRIPPED DURING A ROUTINE INSPECTION. NO PATIENT OR SURGICAL INFORMATION IN AVAILABLE. THIS IS REPORT TWO OF TWO FOR THIS EVENT.
Patient problems: Insufficient Information
MAUDE report 12888408 ↗MalfunctionNov 29, 2021Material Deformation; Material Deformation
IT WAS REPORTED THE TIPS OF TWO SCREW DRIVERS WERE FOUND TO BE STRIPPED DURING A ROUTINE INSPECTION. NO PATIENT OR SURGICAL INFORMATION IN AVAILABLE. THIS IS REPORT ONE OF TWO FOR THIS EVENT.
Patient problems: Insufficient Information
MAUDE report 12888394 ↗MalfunctionNov 23, 2021Material Deformation; Material Deformation
IT WAS REPORTED THE TIP OF THE SCREWDRIVER WAS FOUND TO BE DEFORMED AFTER SURGERY. THERE WAS NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12864694 ↗
Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices
- EXPEDIUM 179702000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL439 reports
- EXPEDIUM VERSE 199721001Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL227 reports
- XIA 3 48230000Stryker Corporation · APPLIANCE, FIXATION, SPINAL INTERLAMINAL193 reports
- VIPER 186715000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL1 death167 reports
- NA 04.614.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL80 reports
- NA 04.615.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- EXPEDIUM VIPER 196789480Depuy Spine, LLC · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- Arsenal Deformity Spinal Fixation System 47127Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL66 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page