Natural Nail® 47-2484-032-50Recalls & reported problems
Zimmer, Inc. · Class 2 · Implant
All models of Natural Nail (52)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00889024093997), received between Apr 7, 2023 and Jun 15, 2023 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Malposition of Device” (2 reports).
15 other recalls cover devices cleared under the same FDA submission (K082770); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Malposition of Device2
- Failure to Align1
Patient problems
- Malunion of Bone2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJun 15, 2023Malposition of Device; Malposition of Device
UPON RECEIPT OF ADDITIONAL INFORMATION, THIS DEVICE WAS DETERMINED TO BE NOT REPORTABLE. IT WAS REPORTED FROM THE SURGEON THAT THE TIBIAL NAIL WAS INITIALLY IMPLANTED IN AN INCORRECT PLACEMENT WHICH LED TO REVISION OF THE NAIL. THE ASSOCIATED SCREWS ARE NOT IMPLICATED IN THE INCORRECT PLACEMENT AND DID NOT CAUSE OR CONTRIBUTE TO THE REPORTED EVENT.
Patient problems: Malunion of Bone
MAUDE report 17134535 ↗InjuryJun 15, 2023Malposition of Device; Malposition of Device
NO FURTHER EVENT INFORMATION AT TIME OF THIS REPORT.
Patient problems: Malunion of Bone
MAUDE report 17134531 ↗MalfunctionApr 7, 2023Failure to Align; Failure to Align
IT WAS REPORTED THAT THE SURGEON WAS IMPLANTING A PERITROCHANTERIC NAIL, AFTER SPEAKING WITH THE CONTROL TEAM DURING THE OPERATION, IT WAS NOTED THAT THE SCREWS HAD BEEN IMPLANTED ABOVE THE NAIL HOLES. THE NAIL WAS REMOVED AND REPLACED WITH ANOTHER BACK UP PRODUCT DURING THE SAME SURGERY. THE SURGEON IS CONCERNED THAT THE TARGETING INSTRUMENT HAD CONTRIBUTED TO THE FAILURE. NO ADVERSE EVENTS HAVE BEEN REPORTED AS A RESULT OF THE MALFUNCTION. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16701625 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K082770. One submission often covers many models, so check the product and lot details in each record.
- Z-3246-2018 · Jul 30, 2018 · TerminatedZimmer Natural Nail Tibial Nail - Yellow 8.3mm, Model Number 47249534008 Product Usage - The Zimmer Natural Nail System is intended for temporary fracture fixation and stabilizati
- Z-0387-2010 · Jul 1, 2009 · TerminatedZimmer Natural Nail System, Tibial Module, nonsterile, Zimmer, Warsaw, IN; REF 00-2490-005-04. The device is used to drill into the proximal tibia or distal femur for placement of
- Z-0388-2010 · Jul 1, 2009 · TerminatedZimmer Natural Nail System, Connecting Knob, nonsterile, Zimmer, Warsaw, IN; REF 00-2490-000-11. The device is used to drill into the proximal tibia or distal femur for placement
- Z-0381-2010 · Jul 1, 2009 · TerminatedZimmer Natural Nail System, Antegrade Femoral Targeting Guide Handle, small, 0.512" wide, nonsterile, Zimmer, Warsaw, IN; REF 00-2490-008-10. The device is used to drill into the
- Z-0377-2010 · Jul 1, 2009 · TerminatedZimmer Natural Nail System, Antegrade Femoral Targeting Guide Handle, standard, 0.512" wide, nonsterile, Zimmer, Warsaw, IN; REF 00-2490-001-10. The device is used to drill into t
- Z-0379-2010 · Jul 1, 2009 · TerminatedZimmer Natural Nail System, Tibial Targeting Guide Handle, tall, nonsterile, Zimmer, Warsaw, IN; REF 00-2490-005-03. The device is used to drill into the proximal tibia or distal
- Z-0378-2010 · Jul 1, 2009 · TerminatedZimmer Natural Nail System, Tibial Targeting Guide Handle, small, nonsterile, Zimmer, Warsaw, IN; REF 00-2490-005-00. The device is used to drill into the proximal tibia or distal
- Z-0386-2010 · Jul 1, 2009 · TerminatedZimmer Natural Nail System, Antegrade Femoral Recon Module, small, nonsterile, Zimmer, Warsaw, IN; REF 00-2490-008-08. The device is used to drill into the proximal tibia or dista
- Z-0380-2010 · Jul 1, 2009 · TerminatedZimmer Natural Nail System, Antegrade Femoral Targeting Guide Handle, small, nonsterile, Zimmer, Warsaw, IN; REF 00-2490-008-00. The device is used to drill into the proximal tibi
- Z-0383-2010 · Jul 1, 2009 · TerminatedZimmer Natural Nail System, Antegrade Femoral Recon Module, standard, nonsterile, Zimmer, Warsaw, IN; REF 00-2490-001-08. The device is used to drill into the proximal tibia or di
- Z-0382-2010 · Jul 1, 2009 · TerminatedZimmer Natural Nail System, Antegrade Femoral Interlock Module, standard, nonsterile, Zimmer, Warsaw, IN; REF 00-2490-001-03. The device is used to drill into the proximal tibia o
- Z-0384-2010 · Jul 1, 2009 · TerminatedZimmer Natural Nail System, Antegrade Femoral interlock Module, tall, nonsterile, Zimmer, Warsaw, IN; REF 00-2490-001-53. The device is used to drill into the proximal tibia or di
- Z-0375-2010 · Jul 1, 2009 · TerminatedZimmer Natural Nail System, Antegrade Femoral Targeting Guide Handle, standard, nonsterile, Zimmer, Warsaw, IN; REF 00-2490-001-00. The device is used to drill into the proximal t
- Z-0385-2010 · Jul 1, 2009 · TerminatedZimmer Natural Nail System, Antegrade Femoral Interlock Module, small, nonsterile, Zimmer, Warsaw, IN; REF 00-2490-008-03. The device is used to drill into the proximal tibia or d
- Z-0376-2010 · Jul 1, 2009 · TerminatedZimmer Natural Nail System, Antegrade Femoral Targeting Guide Handle, tall, nonsterile, Zimmer, Warsaw, IN; REF 00-2490-001-05. The device is used to drill into the proximal tibia
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.