- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00889024199149), received between Oct 10, 2019 and Jun 11, 2025 — classified as 10 injuries.
The most frequently reported specific device problem is “Fracture” (4 reports).
4 other recalls cover devices cleared under the same FDA submission (K013385, K173057); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture4
- Unstable3
- Detachment of Device or Device Component2
- Adverse Event Without Identified Device or Use Problem1
- Device Dislodged or Dislocated1
- Insufficient Device Problem Information1
- Loosening of Implant Not Related to Bone-Ingrowth1
- Material Erosion1
Patient problems
- Failure of Implant5
- Joint Dislocation3
- Pain3
- Ambulation or Postural Difficulties2
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions2
- Scar Tissue2
- Inflammation1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryJun 11, 2025Fracture; Unstable; Fracture; Unstable
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Ambulation or Postural Difficulties; Failure of Implant; Pain; Rash
MAUDE report 22183621 ↗InjuryOct 16, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT UNDERWENT A LEFT KNEE REVISION APPROXIMATELY FIVE MONTHS POST-IMPLANTATION DUE TO STIFFNESS. THE FEMORAL AND INSERT WERE REMOVED AND REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20465761 ↗InjuryApr 2, 2024Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT UNDERWENT A LEFT KNEE KNEE REVISION APPROXIMATELY FIVE YEARS POST-IMPLANTATION DUE TO IMPLANT FRACTURE. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19020025 ↗InjuryOct 10, 2023Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant; Insufficient Information
MAUDE report 17906081 ↗InjurySep 14, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT A PATIENT UNDERWENT A REVISION TOTAL KNEE ARTHROPLASTY. APPROXIMATELY ONE YEAR AND NINE MONTHS POST-IMPLANTATION, THE PATIENT BEGAN TO EXPERIENCE PAIN, INSTABILITY, IMMOBILITY AND MODERATE JOINT EFFUSION. DIAGNOSTIC IMAGES REVEALED THAT THE LOCKING BAR COMPONENT OF THE ARTICULAR SURFACE HAD FRACTURED. SUBSEQUENTLY, THE PATIENT UNDERWENT REVISION SURGERY. DURING THE SURGERY, THE KNEE JOINT WAS FOUND TO BE UNCOUPLED FROM THE CONNECTOR DUE TO BREAKAGE OF THE LOCKING BOLT AS WELL AS DORSAL DISLOCATION OF THE ARTICULAR SURFACE AND LOOSENING OF THE CEMENTED FEMORAL COMPONENT. THE FEMORAL COMPONENTS AND ARTICULAR SURFACE WERE REVISED WITHOUT COMPLICATION. DUE DILIGENCE IS COMPLETE AS MULTIPLE ATTEMPTS HAVE BEEN MADE; ALL AVAILABLE INFORMATION HAS BEEN REPORTED.
Patient problems: Ambulation or Postural Difficulties; Failure of Implant; Joint Dislocation; Joint effusion; Pain
MAUDE report 17745658 ↗InjuryAug 9, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant
MAUDE report 17490381 ↗InjuryMar 2, 2022Unstable; Detachment of Device or Device Component; Migration; Unstable; Detachment of Device or Device Component; Migration
IT WAS REPORTED THAT THE PATIENT UNDERWENT A KNEE ARTHROPLASTY REVISION TO ADDRESS THE FEMORAL SCREW BACKING OUT APPROXIMATELY TEN (10) MONTHS POST-OPERATIVELY AND REPAIR A QUARTER-SIZE HOLE IN THE QUADRICEPS TENDON. THE PATIENT COMPLETED SEVEN (7) WEEKS OF STABILIZATION WITH A FULL-LENGTH LEG CAST AND THEN RECEIVED A FULL LEG BRACE RESTRICTING TO 60 DEGREES AND PHYSICAL THERAPY. ATTEMPTS HAVE BEEN MADE, HOWEVER, NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant; Inflammation; Joint Dislocation; Muscle/Tendon Damage; Scar Tissue; Swelling/ Edema; Synovitis
MAUDE report 13655963 ↗InjuryAug 20, 2021Unstable; Insufficient Device Problem Information; Unstable; Insufficient Device Problem Information
IT WAS REPORTED THAT A PATIENT WAS REVISED APPROXIMATELY FOUR YEARS AND TEN MONTHS¿ POST IMPLANTATION DUE TO UNKNOWN REASONS. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE INFORMATION IS AVAILABLE AT THIS TIME.
Patient problems: Insufficient Information; Joint Laxity
MAUDE report 12348756 ↗InjuryJul 20, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED A PATIENT WAS REVISED DUE TO HINGE POST WAS DISLOCATED FROM ROTATING HINGE KNEE FEMORAL HINGE AND ARTICULAR SURFACE WAS DISLOCATED FROM TIBIAL COMPONENT. ATTEMPTS HAVE BEEN MADE, BUT THERE IS NO ADDITIONAL INFORMATION AT THIS TIME.
Patient problems: Joint Dislocation
MAUDE report 12196293 ↗InjuryOct 10, 2019Material Erosion; Fracture; Material Erosion; Fracture
IT WAS REPORTED THAT A PATIENT WAS REVISED APPROXIMATELY TWO AND HALF YEARS POST IMPLANTATION DUE TO IMPLANT FRACTURE. PATIENT EXPERIENCED LEFT KNEE PAIN AND DIFFICULTY AMBULATING. LEFT KNEE REVISION PERFORMED, NOTED THE HINGE HAD FRACTURED AND FRAGMENT HAD MIGRATED, ALONG WITH POLY FRACTURE. PATIENT HAD TENDON TEAR REPAIRED WITH ANCHORS. THERE IS NO ADDITIONAL INFORMATION AT THIS TIME.
Patient problems: Limited Mobility Of The Implanted Joint; Loss of Range of Motion; No Information; Pain; Scar Tissue; Swelling; Tissue Damage
MAUDE report 9176594 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K013385, K173057. One submission often covers many models, so check the product and lot details in each record.
- Z-2092-2023 · May 19, 2023 · TerminatedNexGen Precoat Stemmed Tibial Plate Size 5
- Z-0321-2020 · Oct 10, 2019 · TerminatedVarious Knee Systems for knee joint arthroplasty, knee prosthesis, Item Nos. 00522000100 00522000200 00522000300 00522000400 00522006304 00522006504 00522006600 00522006800
- Z-1447-04 · Aug 3, 2004 · TerminatedZimmer brand NEXGEN Complete Knee Solution Rotating Hinge Knee Tibial Component, precoat, nonmodular, size 2, sterile, for cemented use only; Cat. No. 5880-02-02.
- Z-1448-04 · Aug 3, 2004 · TerminatedZimmer brand NEXGEN Complete Knee Solution Rotating Hinge Knee Tibial Component, precoat, nonmodular, size 3, sterile, for cemented use only; Cat. No. 5880-03-02.
Other PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER devices
- ELEOS™ 25001208EOnkos Surgical, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER3 deaths193 reports
- ELEOS™ THSMWRS01MOnkos Surgical, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER190 reports
- "UNITED" USTAR II Hip System 1115-1413United Orthopedic Corp. · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer3 recalls3 reports
- "UNITED" USTAR II Hip System 1115-1213United Orthopedic Corp. · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer3 recalls0 reports
- "UNITED" USTAR II Hip System 1115-1417United Orthopedic Corp. · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer3 recalls0 reports
- "UNITED" USTAR II Hip System 1115-1217United Orthopedic Corp. · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer3 recalls0 reports
- Endo-Model Modular Knee Prosthesis System 15-3816/12WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls6 reports
- Endo-Model Modular Knee Prosthesis System 15-3816/11WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page